- Store gentleMACS Octo Cooler for at least 4 hours in upright position at –20 °C. Alternatively place gentleMACS Dissociator in a cold room.
- Use fresh, frozen or OCT-embedded tissue (fixed tissue is not suitable).
- The recommended minimum amout of tissue is 10 mg.
- Recommended: use low protein-binding consumables.
- Pre-cool centrifuge to +4 °C.
Nuclei extraction and enrichment
This application protocol involves the automated extraction of nuclei from the tissue using the gentleMACS® Dissociator, followed by magnetic enrichment using MACS® Technology in combination with the Anti-Nucleus MicroBeads. The purified nuclei can then undergo further processing for applications such as single-nucleus RNA sequencing or ATAC sequencing. This protocol has been validated using fresh or frozen tissue samples from humans and mice. However, the Anti-Nucleus MicroBeads used in this protocol are species-agnostic and can technically be used for any eukaryotic species.
▲ Note: Please note that a different extraction approach and buffer may be required for plant or insect samples.
Compatibility with Anti-Nucleus MicroBeads for non-validated species cannot be guaranteed.
Protocol
Nuclei extraction and enrichment
5 min
Buffer preparation
5 min
Run program on gentleMACS
5 min
Filtering nuclei suspension
5 min
Centrifugation
5 min
Resuspend and count
15 min
Labeling and column preparation
5 min
Elute purified nuclei and count
Preparation of nuclei separation buffer for resuspension
Select the required amount of nuclei separation buffer for resuspension based on the tissue input.
Table 1: Required amount of nuclei separation buffer for resuspension based on the tissue input.
| Tissue input (mg) | Volume of nuclei separation buffer (mL) |
| ≤ 50 | 1.5 |
| 51 – 100 | 3.0 |
| 101 – 200 | 6.0 |
| Mouse brain hemisphere | 8.0 |
▲Note: When using the cells in suspension, it is recommended to spin down the cells and snap freeze the cell pellet in liquid nitrogen before proceeding with the nuclei extraction on the gentleMACS Dissociator. Assume the tissue input for this pellet as 50 mg.
Prepare the required amount of nuclei separation buffer according to the tissue input, as mentioned in the table below, to achieve a final concentration of 0.04% Nuclei Extraction Buffer.
Table 2: Composition of Nuclei Extraction Buffer, PBS, and MACS® BSA Stock to prepare the nuclei separation buffer.
| Volume of nuclei separation buffer (mL) | Nuclei Extraction Buffer (mL) | PBS (pH 7.2) (mL) | MACS BSA Stock (10%) (µL) |
| 1.5 | 0.21 | 1.29 | 6 |
| 3.0 | 0.42 | 2.58 | 12 |
| 6.0 | 0.84 | 5.16 | 24 |
| 8.0 | 1.12 | 6.88 | 32 |
- Always keep the prepared buffer on ice.
- Add RNase inhibitor (0.2 U/µL), if RNA is to be preserved.
- RNase inhibitor can be excluded while performing DNA-based downstream applications.
Preparation of nuclei separation buffer for MACS Separation
Prepare 10.5 mL nuclei separation buffer for the enrichment of 106 total nuclei by combining the components below to achieve a final concentration of 0.04% Nuclei Extraction Buffer in PBS (pH 7.2):
Table 3: Preparation of nuclei separation buffer
| Nuclei Extraction Buffer (mL) | PBS (pH 7.2) (mL) | MACS BSA Stock Solution (10% BSA) (µL) |
| 1.5 | 9.0 | 42 |
▲Note: Prepare nuclei separation buffer for MACS Separation by diluting MACS BSA Stock Solution (10% BSA) (0.04% final concentration) and Nuclei Extraction Buffer 1:7 (14% final concentration) in phosphate-buffered saline (PBS), pH 7.2.
For example, for separation of 106 nuclei, prepare 10.5 mL nuclei separation buffer by adding 1.5 mL of Nuclei Extraction Buffer to 9 mL PBS and MACS BSA Stock Solution (10% BSA). In case of RNAbased downstream applications, add RNase inhibitor (0.2 U/μL final concentration). Keep buffer cold (+2 to +8 °C). Degas buffer before use, as air bubbles could block the column.
▲Note: Prepare appropriate volume of nuclei separation buffer for resuspension by diluting MACS BSA Stock Solution 1:250 (0.04% final concentration) and Nuclei Extraction Buffer 1:7 (14% final concentration) in phosphate-buffered saline (PBS), pH 7.2 as indicated in table 2.
- Always keep the prepared buffer on ice
- Add RNase inhibitor (0.2 U/μL) if RNA is to be preserved.
- RNase inhibitor can be excluded while performing DNA-based downstream applications.
2.1 Nuclei extraction and labeling
▲ Optimal extraction performance depends on the starting material (tissue type, and the amount of starting material). Downstream application may require further optimization of the protocol.
▲ Work fast until the tissue is dissociated and keep the samples on ice after each step.
▲ Volumes given below are for up to 200 mg tissue per extraction.
- Add 2 mL ice-cold Nuclei Extraction Buffer to each pre-cooled gentleMACS C Tube.
- Transfer tissue pieces to the gentleMACS C Tube containing nuclei extraction buffer and directly proceed the following steps until the tissues are dissociated.
▲ Note: Do not allow the frozen sample to be thawed before the dissociation as endogenous RNase might degrade RNA. - Close the gentleMACS C Tube and place it on the gentleMACS Dissociator.
- Place the frozen gentleMACS Octo Cooler on top of the C Tube.
▲ Note: Please check the instrument's compatibility with the gentleMACS Octo Cooler in advance. Alternatively, place the instrument in a cold room (+2 °C to +8 °C). - Run gentleMACS Program 4C_nuclei_1 on the gentleMACS Dissociator.
- After termination of the program, immediately remove the gentleMACS Octo Cooler, detach C Tube from the gentleMACS Dissociator and place the C Tube immediately on ice.
▲ Note: Using low protein binding tubes (or 15 mL tubes) from here onwards may increase the nuclei recovery. - Apply nuclei suspension to a MACS SmartStrainer (70 µm) placed on a 15 mL tube.
▲ Note: For brain samples, use a MACS SmartStrainer (100 µm). - Wash MACS SmartStrainer with 2 mL ice-cold Nuclei Extraction Buffer.
▲ Note: Alternatively, for maximum cell recovery, rinse the used C Tube with ice-cold Nuclei Extraction Buffer before adding to the MACS SmartStrainer. - Discard the MACS SmartStrainer and centrifuge the nuclei suspension at 300 × g at +4 °C for 5 minutes. Carefully aspirate the supernatant completely.
- Resuspend nuclei pellet with ice-cold nuclei separation buffer for resuspension by slowly and gently pipetting the sample up and down for about 10 times.
- Apply nuclei suspension to the MACS SmartStrainer (30 μm) placed on a 15 mL tube.
- Collect nuclei suspension and determine nucleus number.
▲ Note: If the nucleus concentration is higher than 2.22 × 106 nuclei/mL, centrifugation step may be skipped. Measure the volume that contains 1 × 106 nuclei, and add nuclei separation buffer to reach a total volume of 450 µL. Proceed to step 14. - Centrifuge nucleus suspension at 300 × g for 5 minutes. Pipette off supernatant completely.
- Resuspend nuclei pellet in 450 µL of nuclei separation buffer per 106 total nuclei.
- Add 50 µL of Anti-Nucleus MicroBeads per 106 total nuclei.
- Mix well and incubate for 15 minutes in the refrigerator (+2 to +8 °C).
- Add 2 mL nuclei separation buffer (independent from nucleus sample input) and proceed to magnetic separation.
2.2 Magnetic separation
▲ Choose LS Columns and an appropriate MACS Separator.
Alternatively, use Multi-24 Column Blocks with the MultiMACS™ Cell24 Separator or MultiMACS X. Please refer to the MultiMACS Cell24 Separator Plus or the MultiMACS X user manual, respectively, for detailed information on using the Multi-24 Column Blocks for magnetic separation and elution procedure
▲ When using MultiMACS Cell24 Separator Plus for high-throughput nuclei separation, select program Possel_2 to collect the negative fraction as a control, followed by the positive fraction. Select program Possel_1 to isolate only the positive fraction.
▲ When using the autoMACS® NEO Separator for automated nuclei separation, replace the autoMACS Running Buffer in the system with the nuclei separation buffer before use. Please note that a higher volume of nuclei separation buffer is required compared to manual separation.
▲ Always wait until the column reservoir is empty before proceeding to the next step.
- Place LS Column in the magnetic field of a suitable MACS Separator. For details, refer to the respective MACS Column data sheet
- Prepare column by rinsing with 3 mL of nuclei separation buffer.
- Apply nuclei suspension onto the column.
▲ Note: (Optional) Collect flow-through containing debris.
▲ Note: Since the volume of the buffer should be kept low, add the nuclei suspension directly to the center of the column reservoir. - Wash the column two times with 1 mL of nuclei separation buffer.
▲ Note: (Optional) Collect the debris that passes through and combine with the flow-through from step 3.
▲ Note: Perform washing steps by adding buffer aliquots as soon as the column reservoir is empty.
▲ Note: Use buffer to rinse the inner column walls. - Remove column from the separator and place it on a suitable collection tube.
- Pipette the appropriate amount of nuclei separation buffer onto the column depending on nuclei sample input. Immediately flush out the magnetically labeled nuclei by firmly pushing the plunger into the column.
Nuclei sample input Nuclei separation buffer
(mL)≤ 4 × 106 1 ≤ 107 2.5 ≤ 2 × 107 5 - Process immediately with downstream application e.g., single nucleus RNA seq, ATAC-Seq, or nuclei sorting.
2.3 (Optional) Labeling for microscopic or flow cytometric analysis of nuclei samples
- Add 7-AAD Staining Solution (final concentration of 0.525 µg/mL), DAPI Staining Solution (final concentration of 0.25 µg/mL), or DRAQ5TM (10 µM) Staining Solution to a small fraction of the nuclei suspension. Mix by inverting it gently.
- Incubate at +4 °C for 5 minutes.
- Load the sample on a hemocytometer or on a flow cytometer, e.g., MACSQuant® Analyzer.
- Analyze sample according to manufacturer’s recommendation.
18 products found
|
For research use only
For research use only
For research use only
For research use only
25 tubes (sterile, single-packed)
(excl. tax)
For research use only
4 pieces
(excl. tax)
For research use only
for 25 extractions
(excl. tax)
For research use only
(excl. tax)
For research use only
6× 75 mL
(excl. tax)
For research use only
25 columns
(excl. tax)
For research use only
1 piece
(excl. tax)
For research use only
1 piece
(excl. tax)
For research use only
1 piece
(excl. tax)
For research use only
For research use only
For research use only
1 unit
For research use only
EUR 111,00
(excl. tax)
For research use only
1 mL
(excl. tax)
For research use only
1 mL
(excl. tax)
For research use only
200 µL
(excl. tax)
For research use only
8 pieces
(excl. tax)