- Always prepare reagents freshly.
Multicolor flow cytometric analysis of granulocytes from mouse spleen
This application protocol describes the flow cytometric analysis of granulocytes after spleen dissociation from healthy C57BL/6 mice. Viable single cells from mouse spleens are easily obtained using the gentleMACS® Technology. For downstream flow cytometric analysis of different granulocyte populations such as neutrophils, basophils, and eosinophils, two validated multicolor flow cytometry panels were designed using REAfinity™ Recombinant Antibodies and Viobility™ Fixable Dyes. In addition, optimized gating strategies are provided for the analysis of granuloycte populations from mouse spleen.
Protocol
Multicolor flow cytometric analysis of granulocytes from mouse spleen
5 min
Dissociation of mouse spleen
50 min
Staining of granulocytes
Flow cytometric analysis
This protocol includes detailed instructions for preparing necessary reagents and materials and then carrying out each step.
Preparation of the PEB buffer
PEB buffer: Prepare a solution containing phosphate-buffered saline (PBS), pH 7.2, 0.5% bovine serum albumin (BSA), and 2 mM EDTA by diluting MACS® BSA Stock Solution (# 130-091-376) 1:20 with autoMACS® Rinsing Solution (# 130-091-222). Keep buffer cold (2–8 °C). Always use freshly prepared buffer. Do not use autoMACS Running Buffer or MACSQuant® Running Buffer as they contain a small amount of sodium azide that could affect the results.
- For details on the use of gentleMACS Dissociators, refer to the gentleMACS Dissociator user manuals.
- To obtain the results shown in this protocol, one spleen per C Tube has been dissociated.
- Transfer mouse spleens into the gentleMACS C Tube containing the following amount of PEB buffer:
1–2 mouse spleens: 3 mL
3–4 mouse spleens: 6 mL
5–6 mouse spleens: 9 mL - Tightly close the C Tube and attach it upside down onto the sleeve of the gentleMACS Dissociator.
▲ Note: Close the C Tube tightly beyond the first resistance.
▲ Note: Ensure that the sample material is located in the area of the rotor/stator. - Choose and run one of the following gentleMACS Programs:
1–2 mouse spleens: m_spleen_01
3–6 mouse spleens: m_spleen_04 - After termination of the program, detach the C Tube from the gentleMACS Dissociator.
- (Optional) Perform a short centrifugation step to collect the sample material at the bottom of the tube.
- Resuspend sample and apply the cell suspension to a MACS SmartStrainer (70 μm) placed on a 15 mL tube (1–2 mouse spleens per C Tube) or to a 50 mL tube (3–6 mouse spleens per C Tube).
▲ Note: Moisten the MACS SmartStrainer with 2 mL buffer before use. - Wash the filter with 5 mL of PEB buffer.
- Discard the filter and centrifuge cell suspension at 300×g for 10 minutes at room temperature. Aspirate supernatant completely.
- Resuspend cells in PEB buffer to an appropriate volume and proceed to flow cytometry staining.
Table 1: Configuration of the panel to identify neutrophils and eosinophils
| Laser | Antibody specificity | Fluorochrome | Clone |
| Violet | CD11c | VioBlue® | REA592 |
| Viobility 405/520 Fixable Dye | |||
| Blue | Ly-6G | FITC | REA526 |
| CD193 (CCR3) | PE | REA122 | |
| CD3ε* | PE-Vio® 615 | REA606 | |
| CD19* | PE-Vio 615 | REA747 | |
| CD335 (NKp46)* | PE-Vio 615 | REA815 | |
| CD11c | PE-Vio 770 | REA754 | |
| Red | Siglec-F | APC | REA798 |
| CD45 | APC-Vio 770 | REA737 |
* Exclusion channel containing lineage markers to exclude T cells (CD3ε), B cells (CD19), and NK cells (CD335 (NKp46)) from the analysis.
Table 2: Configuration of the panel to identify basophils
| Laser | Antibody specificity | Fluorochrome | Clone |
| Violet | CD11b | VioBlue | REA592 |
| Viobility 405/520 Fixable Dye | |||
| Blue | FcεRIα | FITC | MAR-1 |
| CD117 | PE | REA791 | |
| CD3ε* | PE-Vio 615 | REA606 | |
| CD19* | PE-Vio 615 | REA747 | |
| CD335 (NKp46)* | PE-Vio 615 | REA815 | |
| CD200R3 | PE | REA128 | |
| Red | CD49b | APC | REA541 |
| CD45 | APC-Vio 770 | REA737 |
* Exclusion channel containing lineage markers to exclude T cells (CD3ε), B cells (CD19), and NK cells (CD335 (NKp46)) from the analysis.
Labeling of cells with Viobility 405/520 Fixable Dye
- Volumes given below are for up to 107 nucleated cells. When working with fewer than 107 cells, use the same volumes as indicated. When working with higher cell numbers, scale up all reagent volumes and total volumes accordingly (e.g. for 2×107 nucleated cells, use twice the volume of all indicated reagent volumes and total volumes).
- Determine cell number.
- Wash cells in 1 mL of 1× PBS and centrifuge at 300×g for 10 minutes. Aspirate supernatant completely.
- Resuspend cells in 100 μL of 1× PBS.
- Add 1 µL of Viobility 405/520 Fixable Dye.
- Mix well and incubate for 15 minutes in the dark at room temperature.
▲ Note: Higher temperatures and/or longer incubation times may lead to non-specific cell labeling. Working on ice requires increased incubation times. - Wash cells by adding 1 mL PEB buffer and centrifuge at 300×g for 10 minutes. Aspirate supernatant completely.
- (Optional) Repeat step 6.
- Resuspend cells in an appropriate volume of PEB buffer and proceed with surface staining.
Surface staining
- To obtain the results shown in this protocol a total of 1×106 splenocytes has been stained in a total volume of 100 µL.
- Determine cell number and divide cell suspension equally into two tubes.
- Centrifuge cell suspension at 300×g for 10 minutes. Aspirate supernatant completely.
- Prepare a staining cocktail for each panel with all fluorochrome-conjugated antibodies according to the respective data sheets.
- Add one staining cocktail per tube containing the cell suspension.
▲ Note: Do not add both staining cocktails into the same tube. - Mix well and incubate for 10 minutes in the dark in the refrigerator (2−8 °C).
▲ Note: Higher temperatures and/or longer incubation times may lead to non-specific cell labeling. Working on ice requires increased incubation times. - Wash cells in each tube by adding 1 mL of PEB buffer and centrifuge at 300×g for 10 minutes. Aspirate supernatant completely.
- Resuspend cells in an appropriate volume of PEB buffer and proceed to flow cytometric analysis.
- Prior to acquisition of samples in a flow cytometer, e.g., the MACSQuant® Analyzer, make sure lasers are properly calibrated and compensation of the spillover of fluorochrome emissions has been established in advance using cells or the MACS Comp Bead Kit, anti-REA. Prior to acquisition adjust also forward scatter (FSC)/side scatter (SSC) parameters to properly locate all leukocyte populations in the plots.
- Exclude doublets by using FSC-H/FSC-A parameters.
- Use gated single cells to exclude erythrocytes and debris using the SSC/FSC parameters.
- Use the SSC/FSC gate to exclude dead cells from the analysis using the Viobility 405/520 Fixable Dye/FSC.
- Use gated Viobility 405/520 Fixable Dye–negative (viable) cells to identify granulocyte subsets.