| Optics | |
| Laser excitation | Spatially separated: 405 nm, 40 mW 488 nm, 50 mW 561 nm, 100 mW |
| Emission detectors | FSC: 561/4 nm SSC: 561/4 nm V1: 452/45 nm V2: 525/50 nm Y1: 586/15 nm Y2: 615/20 nm Y3: 661/20 nm Y4: 740 nm LP B1: 525/50 nm B2: 593–650 nm |
| Flow cell dimensions | 200 × 250 μm |
| Fluorescence detectors | Optimized with spectrally matched PMTs for all channels |
| Fluidics | |
| Minimal uptake volume1 | 1 μL (25 μL recommended for volumetric counting applications) |
| Sample flow rate | 25, 50, or 100 µL/min or automate flow rate to maintain 500, 1000, or 2000 events/second |
| Measurement speed2,3 | <25 minutes per 96-well plate (5 μL measurement volume per well) |
| Sample uptake | 1–450 µL |
| Maximal event rate | Up to 15,000 events/second |
| Technical specifications | |
| Footprint | 385.5 × 284.5 mm (15.18×11.2 in) |
| Size (width×depth) Height (adjustable touchscreen) | 669 × 400 mm (26.34×15.75 in) 394–553.5 mm (15.51–21.79 in) |
| Size with miniSampler Plus (width×depth) | 669 × 500 mm (26.34×19.69 in) |
| Weight | 50 kg (110 lbs) |
| Monitor | 15.6" LCD touchscreen |
| Power requirements | 100–240 V~, 50/60 Hz |
| Power consumption (standard use) | 450 W |
| RAM | 8 GB DDR4 (SO-DIMM) |
| Mass storage | 500 GB SSD |
| Ports | 4× USB 2.0 ports, 6× USB 3.0 ports (2 at display), 2× DisplayPort, 2× LAN, DVI, RE-232, Audio |
| Emission sound pressure level at workstation | <61 dB(A) |
| Performance | |
| Sample carry-over2,4 | 0.01% |
| Fluorescence performance | 5-decade logarithmic scale (10-2 to 103), displayed in lin, log, or hlog scales |
| Data management | |
| Operating system | Embedded operating system |
| Measurement parameters | Area, width, height for all parameters, with time and volume |
| Signal processing | >18 bit dynamic range in area with 32 bit floating point signal processing |
| Compensation | Automated, or manual with 8×8 matrix, during or after acquisition |
| Data files | .mqd (proprietary file format), .fcs (2.0, 3.0, 3.1 compatible) |
| 1 At every uptake, an additional excess volume is aspirated by the instrument. The excess volumes are calibration- and process-dependent and do not exceed 10 μL for Fast, Standard, and Extended modes, and 20 μL for Screen mode. 2 Referred value indicates the average of multiple experiments and can differ for individual sample materials. 3 The measurement speed is determined by measuring the time between the movement of the robotic arm into the first measured well, and its movement out of the last measured well. The measurements were carried out at high flow rate in fast mode. 4 For carry-over, full 96-well plates were loaded with 200 μL/well of PBMC suspension at a nominal concentration of 10,000 cells/μL in every other well ("SRC-wells”). Alternating wells are loaded with an equal volume of MACSQuant® Running Buffer ("CO-wells”). The uptake volume was set to 100 μL and measured at medium flow rate in standard mode. The carry-over is defined by sum (CO-singlet count)/sum(SRC-singlet count) ×100%. | |
MACSQuant® VYB Flow Cytometer
Your benchmark for fluorescent protein flow cytometry
Flow cytometry with the MACSQuant Family is defined by performance, automation, compactness, and convenience. Accordingly, the MACSQuant VYB offers a comprehensive suite of automation features for start‑up, maintenance, acquisition, analysis, and shutdown. It also features a uniquely configured optical bench with violet, yellow, and blue lasers. This makes the MACSQuant VYB a perfect fit for detecting fluorescent protein reporters and provides optimized separation of PE and FITC conjugates – ensuring consistent, high‑quality results.
MACSQuantify™ Software: Simplify flow cytometry anytime, anywhere
- Multi-instrument alignment: Easy assay transfer from one instrument to another.
- Workflow automation with Express Modes: Automate every step including sample acquisition, automated gating, analysis, and reporting.
- Automated fluorescence compensation: A straightforward, convenient, and effective way for achieving accurate and reliable results.
- Compliance with 21 CFR Part 11: Streamline data submission to regulatory agencies.
- Performance and QC tracking tool: Ensure reliable instrument performance through longitudinal monitoring.
Our services
- Instrument service: Ensuring maximum performance, uptime, and regulatory compliance.
- Custom application service: Automate your custom application and enhance the consistency of your assays through Custom Express Modes.
- Automation services: Custom network and hardware integration for increased standardization and throughput.
Support at your fingertips
- MACSQuant Live Support: Get instant remote support at the click of a button to keep your experiments running.
- Technical and field application support: Expert support throughout your project from tailored workflows and assay design to troubleshooting.
- Hands-on training courses at Miltenyi University: Designed to support every experience level – from foundational knowledge to advanced instrument operation.
MACSQuant VYB Flow Cytometer
Your benchmark for fluorescent protein flow cytometry.
Catalog no.: 130-096-116. For research use only.