Generation of IG heavy- and light-chain cDNA libraries using the BCR RNA Profiling Kit, human


This application protocol describes the generation of IG heavy- and light‑chain cDNA libraries using a combination of highly sensitive multiplex PCR and Unique Molecular Identifiers (UMIs) with the BCR RNA Profiling Kit, human. UMIs are introduced along with first‑strand cDNA synthesis and enable error‑free reconstruction of full‑length Immunoglobulin (IG) heavy- and light‑chain sequences with hypermutations (except in the FR1 region). In addition, UMIs enable for quantification of template cDNA molecules, control of input bottlenecks, and accurate normalization of samples for comparative analysis. Gene‑specific primers targeting IG heavy‑chain constant regions facilitate discrimination of all IGH isotypes, including IgM, IgD, IgG3, IgG1, IgA1, IgG2, IgG4, IgE, and IgA2. For details regarding the required materials and instruments, please go to the Materials tab.

Protocol

The BCR RNA Profiling Kit, human supports an RNA input equivalent to 100 to 500,000 sorted or purified memory/naive B cells, 10 to 1,000 plasma cells/plasmablasts, from Peripheral blood mononuclear cells (PBMCs), or from B cell–containing tissues, and produces indexed, ready-to-sequence libraries for Illumina®‑based sequencing. For sorted PBMCs and B Cells, this protocol also works directly from cells, allowing users to skip prior RNA isolation (Figure 1).

General recommendations to prevent contamination

General recommendations, such as using lab coats, gloves, tips with aerosol filters, certified RNAse/DNAse-free reagents, and decontamination of the workspacebefore working with RNA (e.g. using AppliChem RNase-ExitusPlus™), to minimize the risk of RNA degradation and contamination should be followed.
Performing non-template control reactions is strongly recommended for each experiment. To prevent cross-sample contamination, library preparation should be performed in two physically separated workspaces (PCR boxes): one for pre–PCR and one for post–PCR, located in different rooms. Avoid transferring anything (reagents, pipettes, racks, etc.) from the post–PCR to the pre–PCR workspace. In the case of possible contamination, use decontaminating procedures with special reagents (e.g. using AppliChem DNA/RNA-ExitusPlus).
Cell isolation, RNA purification, cDNA synthesis, first- and second-round PCR master mix preparations should be performed in the pre–PCR workspace. After the first PCR amplification, the tubes should be transferred to the post–PCR workspace to perform all remaining steps except PCR master mix preparation.

Starting material

As starting material, isolated RNA from sorted/isolated B cells, PBMCs, or B cell–containing tissues can be used. Furthermore, it is possible to start directly from intact PBMCs or B cells.

When starting directly from cells, no RNA isolation is needed. Cells should be pelleted by centrifugation and kept on ice during processing. Direct-from-cells procedure was tested for fresh and frozen samples containing up to 20,000 PBMCs.

When starting with RNA, it is necessary to use a Qubit® fluorimeter or other fluorescent-based methods to determine the RNA concentration. High RNA quality is critical for efficient library preparation. Residual traces of proteins, salts, or other contaminants may decrease the efficiency of the enzymatic activity necessary for optimal targeted enrichment. Refer to section "5. Appendix A: Preparing starting material" for more details about recommendations regarding RNA quality for use as input to the protocol. gDNA can significantly affect the cDNA synthesis reaction and subsequent PCR amplification. It is strongly recommended to perform DNase treatment during RNA extraction.

For quantitative IG repertoire profiling, we recommend separating B cell populations with different IG RNA transcription levels (i.e. plasma cells/plasmablasts and memory/naive B cells) by magnetic-activated cell sorting.

In appendix A provides recommendations for RNA Isolation under the following conditions:

  • RNA isolation for large samples (>50,000 cells)
  • RNA isolation for small samples (i.e. <50,000 cells)

If further guidance is needed, some frequently occurring situations are described in appendix A and help to choose the appropriate strategy for RNA isolation from:

  • PBMCs
  • 100 to 50,000 sorted or purified B cells
  • more than 50,000 sorted or purified B cells
  • B cell–containing tissues

It is also possible to use TRIzolTM reagent to isolate RNA.

Please refer to appendix B for a detailed protocol.

Additional information

Before starting the experiment, it is recommended to read the datasheet of the BCR RNA Profiling Kit, human. Also read the specifications of the necessary materials and instruments that are described on the Materials tab of the protocol webpage.

Download data sheet

BCR RNA Profiling Kit, human

Start of the experiment

  • Perform cDNA synthesis, clean-up after cDNA synthesis, and the PCR1 and PCR2 master mix preparation in a pre–PCR workspace.
  • All samples and reagents from the BCR RNA Profiling Kit Box 1 must be kept on ice during reaction set-up. After use, immediately return enzymes to the freezer.
  • Clean-up steps are performed at room temperature (+15 to +25 °C).
  • Thoroughly mix all components before use.
  • For mixing by pipetting, especially at low volumes, it is advised to set the pipette volume to approximately 80% to avoid creating bubbles.

cDNA synthesis starting from RNA

Before start: Thaw RNA samples on ice. Thaw the BCR cDNA Synthesis Primers briefly at room temperature (+15 to +25 °C), mix primers by vortexing, and spin down. Take out the RT Master Mix from the –20 °C freezer, mix briefly by vortexing, and spin down. Place all components on ice until use.

1. In nuclease-free PCR tubes, mix the following components in a final reaction volume of 12 μL per reaction according to table 1.

Table 1: cDNA Synthesis Primer Mix

ComponentVolume (in μL)
RNA (0.1–500 ng per reaction)1–10
BCR cDNA Synthesis Primers2
Water0–9
Total volume per reaction 12

2. Pre-heat a thermal cycler at +70 °C. Incubate the samples for 2 minutes at +70 °C, followed by 2 minutes at +4 °C.

3. During incubation, prepare the Reverse Transcription mix in a final volume of 8 µL per reaction according to table 2.

Table 2: Reverse Transcription mix

ComponentVolume (in μL)
RT Master Mix4
Water4
Total volume per reaction 8

4. After incubation, add 8 μL of the Reverse Transcription mix to each sample, mix well by pipetting up and down 15 times without creating bubbles. Spin briefly and incubate for 30 minutes at +55 °C (lid temperature set to +60 °C).

5. Proceed with clean-up after cDNA synthesis (see chapter Clean-up after cDNA synthesis).

cDNA synthesis starting directly from cells

The direct‑from‑cells procedure was tested on fresh and frozen samples containing up to 20,000 PBMCs.

Before start: Thaw the BCR cDNA Synthesis Primers briefly at room temperature (+15 to +25 °C). Mix each component thoroughly by pipetting up and down before use, and spin down. Take the RT Master Mix from –20 °C freezer, mix briefly by vortexing, and spin down. Place all components on ice until use.

For working with input of under 1,000 B cells, use cells that have been sorted directly into a tube containing cDNA Synthesis master mix. Skip the pelleting step to avoid the risk of losing the sample and start with step 7. For frozen samples (cell pellets), also start with step 7.

1. Transfer cells into a nuclease-free 1.5 mL tube.

2. Centrifuge at 350 × g for 5 minutes at room temperature (+15 to +25 °C).

3. Wash the cells once with 1 mL PBS without Ca2+, Mg2+, or BSA and repeat centrifugation at 350 × g for 5 minutes.

4. Remove the supernatant by carefully removing 900 µL using a 1 mL pipette tip.

5. Continue to remove the supernatant by carefully removing approximately 90 µL using a 100 or 200 µL pipette tip. Be careful not to touch the side of the tube where the pellet is located.
▲ Note: The remaining PBS will not influence the reaction performance.

6. Place the tube on ice.

7. Prepare the cDNA Synthesis master mix for cells of the following components in a final volume of 16 µL per reaction according to table 3.

Table 3: cDNA Synthesis master mix for cells

ComponentVolume (in μL)
RT Master Mix4
Water5
Lysis Solution5
BCR cDNA Synthesis Primers2
Total volume per reaction16

8. If starting from frozen cell pellets, thaw the samples on ice until no frozen sample material is visible.

▲ Note: If starting with cells sorted directly into the cDNA Synthesis master mix, sort the target cell population straight into the tube containing the cDNA Synthesis master mix. Up to 1,000 cells can be sorted into the master mix volume of 16 µL, provided the total droplet volume does not exceed 4 µL. After sorting, briefly spin down the tube to collect any droplets from the walls of the tube. Then gently pipette up and down 10 times, avoiding bubble formation, and proceed to step 10.

9. Pipette 16 µL of the cDNA Synthesis master mix to the tube containing cell pellet. Mix carefully by pipetting up and down 10 times without creating bubbles.

10. Transfer the samples into the corresponding tube of a PCR strip, keep on ice until all samples are transferred.

11. Centrifuge briefly and incubate the samples for 30 minutes at +55 °C (lid temperature set to +60 °C).

12. Proceed with clean-up after cDNA synthesis

Clean-up after cDNA synthesis

Purify the obtained cDNA synthesis products using AMPure® XP Beads or SPRIselectTM reagent (Beckman CoulterTM) as follows:

Before start: Freshly prepare 80% ethanol and adjust beads to room temperature according to manufacturer’s instructions.

  • Before use, thoroughly vortex the beads until homogenized and no pellet remains at the bottom of the bottle.
  • Add 20 μL of beads to each reaction and mix well by pipetting up and down 10 times without creating bubbles.
  • Incubate the solution for 5 minutes at room temperature (+15 to +25 °C).
  • Place the tubes on the magnetic rack and wait until the solution becomes clear blue.
  • While the tubes are on the magnetic rack, carefully remove the supernatant.
  • Keep the tubes on the magnetic rack and add 200 μL of freshly prepared 80% ethanol. Wait for 30 seconds and remove the ethanol.
  • Remove tubes from the magnetic rack. Briefly spin the tubes down and place them back on the magnetic rack. Remove all the residual ethanol carefully, using a small pipette.
  • Remove tubes from the magnetic rack. Add 21 μL of water to the beads, mix well by pipetting up and down 15 times without creating bubbles, and incubate for 5 minutes at room temperature (+15 to +25 °C).
  • Place the tubes on the magnetic rack and wait for the solution to clear.
  • While the tubes are on the magnetic rack, carefully transfer 20 µL of the supernatant into a new tube.
  • Homogenize beads by vortexing thoroughly, as described in step 1.
  • Add 20 μL beads and mix well by pipetting up and down 10 times without creating bubbles.
  • Incubate the solution for 5 minutes at room temperature (+15 to +25 °C).
  • Place the tubes on the magnetic rack and wait for the solution to clear.
  • While tubes are on the magnetic rack, carefully remove and discard the supernatant.
  • Add 200 μL of freshly prepared 80% ethanol to the tubes, while they remain on the magnetic rack. Wait for 30 seconds and remove the ethanol.
  • Remove tubes from the magnetic rack. Briefly spin the tubes down and place them back on the magnetic rack. Remove all the residual ethanol carefully, using a small pipette.
    ▲ Note: It is important to remove all traces of ethanol at this step as it could inhibit subsequent PCR reactions.
  • Remove tubes from the magnetic rack. Add 21 μL of Elution Buffer to the beads for elution and mix well by pipetting up and down without creating bubbles.
  • Incubate the solution for 5 minutes at room temperature (+15 to +25 °C).
  • Place tubes on the magnetic rack and wait for the solution to clear.
  • Keep tubes on the magnetic rack and carefully transfer 20 μL of the cDNA into a new nuclease-free reaction tube.
    ▲ Note: During the elution steps, beads clumping can significantly reduce the cDNA yield. In case bead clumping is observed, please refer to the chapter Troubleshooting - Decreased cDNA synthesis / yield.

First PCR amplification

Before start: Thaw BCR-H PCR1 and BCR-K/L PCR1 Primers briefly at room temperature (+15 to +25 °C) and PCR Master Mix on ice. Mix each component thoroughly by pipetting up and down before use and spin down. Place all components on ice until use.

▲ Note: To adjust for the 2−3-fold higher BCR-K/L mRNA expression levels, it is recommended to use 15 µL eluate for BCR-H and 5 µL for BCR-K/L reaction as input. All further protocol steps are based on this experimental setup.

▲ Note: Depending on the scope of the experiment, the cDNA input for PCR1 may be adjusted between 0 and 20 µL of cDNA eluate for the BCR‑H and BCR‑K/L reactions, respectively. For example, if only BCR‑H is of interest, the entire 20 µL of cDNA eluate may be used for the BCR‑H PCR1, skipping the BCR‑K/L reaction. Please ensure that the volume of water is adjusted accordingly. However, modifying the cDNA input may require additional optimization of PCR cycle numbers and adaptation of sequencing recommendations.

1. At this point, each cDNA sample must be split into two nuclease-free tubes: 15 µL cDNA for BCR-H in one tube and 5 µL cDNA for BCR-K/L library in another tube.

2. For the BCR-H PCR1 master mix, use a final volume of 35 µL per reaction:

Table 4: BCR-H PCR1 master mix

ComponentVolume (in µL)
Water8
PCR Master Mix25
BCR-H PCR1 Primers2
Total volume per reaction35

3. For the BCR-K/L PCR1 master mix, use a final volume of 45 µL per reaction.

Table 5: BCR-K/L PCR1 master mix

Component Volume (in µL)
Water18
PCR Master Mix25
BCR-K/L PCR1 Primers2
Total volume per reaction45

4. Add the appropriate master mix to the purified cDNA samples:

  • Add 35 μL of BCR-H PCR1 master mix to 15 μL cDNA sample. Mix gently by pipetting 10 times without creating bubbles and centrifuge briefly.
  • Add 45 μL of BCR-K/L PCR1 master mix to 5 μL cDNA sample. Mix gently by pipetting 10 times without creating bubbles and centrifuge briefly.

5. Perform a PCR using the following parameters according to table 6.

Table 6: Thermal cycler condition for the first PCR

Lid temperature Reaction volumeRun time
+105 °C50 µL 
StepTemperatureTime (hh:mm:ss)
1+95 °C00:15:00
2+94 °C00:00:30
3+60 °C00:01:30
4+72 °C00:00:30
Go to step 2, repeat X cycles*
5+68 °C00:04:00
Hold+4 °CHold

* Choose the number of cycles according to the amount of RNA/B cells used as input for the experiment (see table 10).

Clean-up after first PCR

▲ Proceed in the post-PCR workspace.

  • Before use, thoroughly vortex the beads until homogenized and no pellet remains at the bottom of the bottle.
  • Add 40 μL of beads to each reaction and mix well by pipetting up and down 10 times without creating bubbles.
  • Incubate the solution for 5 minutes at room temperature (+15 to +25 °C).
  • Place the tubes on the magnetic rack and wait for the solution to clear.
  • While the tubes are on the magnetic rack, carefully remove the supernatant.
  • Keep the tubes on the magnetic rack and add 200 μL of freshly prepared 80% ethanol. Wait for 30 seconds and remove the ethanol.
  • Repeat the washing step with 80% ethanol without removing the tubes from the magnetic rack.
  • Remove tubes from the magnetic rack. Briefly spin the tubes down and place them back on the magnetic rack. Remove all residual ethanol carefully, using a small pipette.
  • Remove tubes from the magnetic rack. Add 21 μL of Elution Buffer to the beads for elution and mix well by pipetting up and down 15 times without creating bubbles.
  • Incubate the solution for 5 minutes at room temperature (+15 to +25 °C).
  • Place the tubes on the magnetic rack and wait for the solution to clear.
  • Keep tubes on the magnetic rack and carefully transfer 20 μL of the purified first PCR products into new nuclease-free reaction tubes, one for BCR-H and one for BCR-K/L samples.
    ▲ Note: It is important to remove all the traces of ethanol at this step as it could inhibit subsequent PCR reaction.
    ▲ Note: The purified PCR product can be stored up to 1 month at −20 °C.

Second PCR amplification

Before start: Thaw BCR-H PCR2 and BCR-K/L PCR2 Primers briefly at room temperature (+15 to +25 °C) and the PCR Master Mix on ice. Mix each component thoroughly by pipetting up and down before use and spin down. Thaw the DNA/RNA UD Indexes (Illumina) briefly at room temperature (+15 to +25 °C) and spin them down. Place all components on ice until use.

1. Prepare primer dilution for BCR-H PCR2-A and BCR-K/L PCR2-B Primers: For this, calculate the volume of PCR2 primers required to process N number of samples (1 μL of diluted primer mix is required per sample). Prepare at least 10 μL. Both primers should be diluted with water by a dilution ratio of 1:10 (e.g., 2 µL primer + 18 µL water = 20 µL of 1:10 primer dilution). Discard the left-over diluted aliquots after use.

2. Prepare two separate PCR2 master mixes in a final volume of 22 μL per reaction without indexes according to tables 7 and 8.

Table 7: BCR-H PCR2 master mix

ComponentVolume (in μL)
Water8.5
PCR Master Mix12.5
BCR-H PCR2 Primers (1:10 diluted)1
Total volume per reaction 22 µL

Table 8: BCR-K/L PCR2 master mix

Component Volume (in μL)
Water8.5
PCR Master Mix12.5
BCR-K/L PCR2 Primers (1:10 diluted)1
Total volume per reaction22 µL

3. Mix well by pipetting up and down and centrifuge briefly.

4. For each reaction, add 22 μL of corresponding PCR2 master mix to a nuclease-free PCR tube, respectively, for BCR-H and BCR-K/L samples.
▲ Note: Proceed to the post–PCR workspace.

5. Add 2 μL of the purified first PCR product to each of the BCR-H PCR2 master mix and BCR-K/L PCR2 master mix.

6. Add 1 μL of DNA/RNA UD Index (Illumina) to each reaction. Use a unique index for each reaction and record their assignment.

7. Mix by pipetting up and down 10 times without creating bubbles and centrifuge briefly.

8. Perform a PCR with the BCR-H and BCR-K/L samples using the parameters from table 9.

Table 9: Thermal cycler condition for the second PCR

Lid temperature
Reaction volumeRun time
+105 °C25 µL 
StepTemperatureTime (hh:mm:ss)
1+95 °C00:15:00
2+94 °C00:00:30
3+55 °C00:01:30
4+72 °C00:00:30
Go to step 2, repeat X cycles*
5+68 °C00:04:00
Hold+4 °CHold

* Choose the number of cycles according to the amount of RNA/B cells used as input for the experiment (see table 10).

▲ Note: The optimal number of second PCR cycles may vary for different templates, cell types, thermal cyclers, and sample amounts. It is recommended to determine the minimal number of PCR cycles required to obtain a sufficient amount (at least 30–40 ng of each library, specific peak, dependent on initial input concentrations and cycles applied) empirically for each experiment (see table 10 & 11).

▲ Note: Given values are based on an adjustment for the 2–3-fold higher BCR-K/L mRNA expression level and the usage of 15 µL eluate for BCR-H and 5 µL for BCR-K/L reaction as input (recommended in section First PCR amplification).

Furthermore, it is recommended to generate parallel libraries of similar nature (e.g., ten samples of 2,000–20,000 sorted B cells each, BCR-H and BCR-K/L chains) using the same number of PCR cycles and mix the obtained libraries in equal volume proportions for sequencing.

This allows obtaining an even coverage in terms of reads-per-cDNA (reads-per-UMI). For example, the library generated with 20,000 B cells may produce more PCR product than one with 2,000 B cells after the same number of PCR cycles. However, the library from 20,000 cells would also carry proportionally more IG cDNA molecules and thus requires more sequencing reads to achieve a comparable coverage.

Table 10: Recommended PCR cycle numbers for sorted memory/naive B cells or PBMCs.

Estimated count of memory/naive B cells in a sampleAmount of template RNA purified from sorted B cells or PBMCRecommended number of first PCR cycles (total cycles)Recommended number of second PCR cycles (total cycles)Recommended number of sequencing reads
1×102 – 5×1020.1–5 ng2513–16100 reads per input memory/naive B cell (calculated per library based on total input amount)
5×102 – 5×1030.5–50 ng2112–15
5×103 – 5×1045–500 ng1812–15
5×104 – 5×10550–500 ng1810–13

Overview of recommended total PCR cycle numbers for sorted memory/naive B cells or PBMC.

▲ Note: Consider that plasma cells may prominently increase in PBMC after infection or vaccination. Keep in mind that plasma cells may have 100– to 1,000-fold higher levels of IG mRNA expression, which requires appropriate sequencing coverage as described in table 11. For average PBMC samples, please refer table 12 for recommended sequencing depth.

Table 11: Recommended PCR cycle numbers for sorted plasma cells and plasmablasts. Note that plasma cells express 100– to 1,000-fold more IG mRNA compared to other B cell populations.

Number of plasma cells in a sampleRecommended number of first PCR cycles (total cycles)Recommended number of second PCR cycles (total cycles)Recommended number of sequencing reads
10–200 plasma cells2212–1525,000 reads per
input plasma cell
200–2,000 plasma cells2012–15

▲ Note: The given number of sequencing reads represents the recommended coverage that allows for UMI‑based elimination of amplification biases and accumulated errors.

Quality control and pooling

  • Verify quality and concentration of the obtained PCR product by capillary electrophoresis (TapeStation® or Bioanalyzer®).
    The fragment size of the library should have a peak at approximately 530–600 bp for BCR-H and 400–480 bp for BCR-K/L chains respectively, when using D5000 ScreenTapes (TapeStation). Please note that using other screen tapes or BioAnalyzer reagents, the peak range can differ.
  • If multiple libraries were generated, pool the obtained second PCR libraries of similar nature (i.e. libraries generated with similar cell/RNA input amounts) separately for BCR-H and BCR-K/L. To do this, combine equal volume portions from each individual sample after the second PCR.
    If 1–4 libraries are prepared at the same time, the whole reaction volumes of the second PCRs can be pooled. For more than four samples, pool 5–10 µL from each second PCR. The total amount of pooled DNA products should be at least 50 ng.

Library clean-up

Before start: Adjust beads to room temperature (+15 to +25 °C) according to manufacturer’s instructions.
Purify the obtained pooled DNA using bead protocol with 1:0.8 sample-to-bead ratio as follows:

  • Before use, thoroughly vortex the beads until homogenized and no pellet remains at the bottom of the tube.
  • Аdd the volume of beads equal to 0.8 the volume of pooled DNA to each reaction and mix well by pipetting up and down 10 times without creating bubbles. Incubate the solution for 5 minutes at room temperature (+15 to +25 °C).
  • Place the tubes on the magnetic rack and wait for the solution to clear.
  • While the tubes are on the magnetic rack, carefully remove and discard the supernatant.
  • Add 200 µL of freshly prepared 80% ethanol to the tubes, while on the magnetic rack. Wait for 30 seconds and remove the ethanol.
  • Repeat the washing step with 80% ethanol without removing the tubes of the magnetic rack. Remove tubes from the magnetic rack. Briefly spin the tubes down and place them back on the magnetic rack. Remove all the residual ethanol carefully, using a small pipette.
    ▲ Note: It is important to remove all traces of ethanol at this step as it could inhibit subsequent sequencing procedures.
  • Remove tubes from the magnetic rack. Add 21 μL of Elution Buffer for elution and mix well by pipetting up and down without creating bubbles.
  • Incubate the solution for 5 minutes at room temperature (+15 to +25 °C).
  • Place the tubes on the magnetic rack and wait for the solution to clear.
  • Keep tubes on the magnetic rack and carefully transfer 20 μL of purified second PCR products into a new tube, respectively for BCR-H and BCR-K/L samples.

    ▲ Note: When generating libraries with low RNA input amounts, a non-specific band (smaller than target band) can be observed. If this is the case, please perform an additional clean-up (after the final 0.8× library clean-up) with a modified bead-to-sample ratio of 0.7× (see section Library clean-up) to remove/reduce the unspecific band.

Final quality control

  • Verify the quality of the pooled BCR-H and BCR-K/L libraries by analyzing an aliquot by capillary electrophoresis (Bioanalyzer or TapeStation) (see figure 3 for an example of Agilent TapeStation traces). Quantify the obtained purified pools using Qubit fluorimeter and Qubit dsDNA High Sensitivity Assay Kit.
  • BCR-H and BCR-K/L pools can be mixed in equimolar ratio for sequencing, considering that initially ¾ of cDNA was used for BCR-H and ¼ for BCR-K/L library preparations.
    ▲ Note: Purified PCR product can be stored for 1–3 months at −20 °C.

Shop required products for this application protocol

1 product found

|