How CliniMACS® Cell Separation works
Principle of magnetic cell separation
The CliniMACS® Plus Instrument utilizes a reagent consisting of an antibody that specifically binds to a specific surface marker on the target cell. The antibody is conjugated to an iron-containing particle that is only nanometers in size and biocompatible.
The enrichment of target cells or depletion of unwanted cells is accomplished by passing the antibody/magnetically-labeled cell suspension through a magnetic separation column, which is provided as part of a single-use disposable tubing set. Labeled cells are retained within the separation column, while unlabeled cells flow through. Recovery of enriched cells is achieved by removing the magnetic field and eluting the targeted cells into a collection bag.
Tubing Sets for CliniMACS Plus Instrument
Applications