Using the MACS® dPCR LENTIcheck Kit (BaEV) with QX200TM / QX600TM / QX ONE Droplet DigitalTM PCR Instruments


This application protocol describes the detection of the gene encoding for the baboon envelope glycoprotein (BaEV) using digital PCR (dPCR) BIO-RAD® QX200 / QX600 / QX ONE Droplet Digital PCR Instruments in combination with the MACS dPCR LENTIcheck Kit (BaEV), human.
The MACS dPCR LENTIcheck Kit (BaEV), human has been developed for dPCR systems to determine the linearity and sensitivity of the BaEV gene detection. Additional positive and negative controls are included to ensure proper detection of the BaEV gene. No signals should be gained for the viral vector or the transduced cells under investigation.

Protocol

▲ All components of the MACS dPCR LENTIcheck Kit (BaEV), human are ready to use. When MACS dPCR LENTIcheck (BaEV) Control was stored at –20 °C, reheat at +50 °C for 20 minutes. Adjust all reagents to room temperature (+19 to +25 °C) before use.
▲ Before opening, mix the components of the MACS dPCR LENTIcheck Kit (BaEV), human briefly with a vortex mixer before use and perform a short spin.
▲ For more information about the Droplet Digital PCR Instruments, refer to the respective user manual.

Preparation of mastermix

1. Each run requires a master mix for three reactions for the no-template control (NTC) (3 × 17 µL), 5 × 3 reactions for the standard curve (5 × 3 × 17 µL), and 5 × 5 reactions for each sample including the respective controls (5 × 5 × 17 µL) (see table 1, 2, and in tab Materials table 6).

2. Prepare a master mix solution for the MACS dPCR LENTIcheck (BaEV) assay in the DNA-free PCR workstation according to table 1. Mix well and perform a short spin.

Table 1: Master mix for the QX200 / QX600 / QX ONE Droplet Digital PCR Instruments.

ComponentVolume per reaction (µL) Volume per n reactions (µL)Final concentration
MACS dPCR LENTIcheck (BaEV) Primers1n0.9 µM each primer
MACS dPCR LENTIcheck (BaEV) Probe1n0.25 µM/probe
2x ddPCR Supermix for Probes (No dUTP)11n × 111×
MluI restriction endonuclease (20,000 units/mL)0.25n × 0.250.23 units/µL
BclI restriction endonuclease (20,000 units/mL)0.25n × 0.250.23 units/µL
Water3.5n × 3.5 
Total volume w/o DNA17n × 17

3. In the DNA-free PCR workstation, add 17 μL of master mix (see table 1) into a 96-well plate for each of the no-template control (NTC; 3 wells), 5 × 3 wells for the positive control dilutions (Standard (ST) 1–5) and 5 × 5 wells for each test sample including the respective controls (see table 2 and in tab Materials table 6). A loading scheme is shown in table 2.

Table 2: Loading scheme of a 96-well PCR plate for the QX200 / QX600 / QX ONE Droplet Digital PCR Instruments.

 123456789101112
AST 5
5 copies /µL
ST 4
1 × 101 copies /µL
ST 3
1 × 102 copies /µL

ST 2
1 × 103 copies /µL
ST 1
MACS dPCR LENTIcheck (BaEV) Kit Control
1 × 104 copies /µL
NTC Extraction control negative sample 3
BNegative control sample (non-transduced cells) 3
CReplicate 2 – test sample 3 (transduced T cells)
D
DNA extraction negative control test sample 1Negative control sample (non-transduced cells) sample 1Replicate 1 – test sample 1 (transduced cells)Replicate 2 – test sample 1 (transduced cells)DNA extraction positive control sample 1Extraction control negative sample 2Negative control sample (non-transduced cells) 2Replicate 1 – test sample 2 (transduced T cells)Replicate 2 – test sample 2 (transduced T cells)DNA extraction positive control sample 2Replicate 2 – test sample 3 (transduced T cells)DNA extraction positive control sample 3
E
F
G
H

4. Add 5 µL of nuclease-free water to each of the 3 no-template control (NTC) wells of the 96-well plate marked as NTC in table 2. Mix well by pipetting up and down.

5. Add 22 µL of ddPCR Buffer Control for Probes (1:1 mixture of nuclease-free water and ddPCR Buffer Control for Probes) to each of the non-used wells to completely fill the used column of the 96-well plate.