Tips and tricks for tissue clearing and sample preparation
Tissue clearing allows for 3D visualization and imaging of entire opaque tissues. Numerous tissue clearing protocols were published that can be attributed to the following main classes: aqueous-based, hydrogel-based, and organic solvent-based protocols. Organic solvent-based protocols are considered as fast and efficient. Miltenyi Biotec’s MACS® Clearing Kit was developed as a non-toxic variant of organic solvent-based tissue clearing. To support users of organic tissue clearing we compiled this list of tips and tricks. Covering common problems and pitfalls, it describes potential causes and recommends solutions to overcome these issues.
To access protocols click here.
Content
- Tissue preparation and fixation
- Labeling
- Tissue clearing
- Imaging
Tissue preparation and fixation
Despite undergoing perfusion fixation, the persistence of a non-white or bright appearance in tissues like liver can be due to incomplete fixation (under-fixation). See the potential causes and recommended solutions:
Incomplete fixation (under-fixation)
1. Cause: Insufficient perfusion or missing post-fixation immersion perfusion.
Solution:
- If under-fixation is noticed early enough, an additional fixation step in the very beginning could help.
- No salvaging is possible after an organ is cleared; improve the fixation conditions for the next experiment.
2. Cause: Wrong fixative concentration
Solution:
- Use recommended concentration, e.g., 4% PFA.
3. Cause: Fixative is expired
Solution:
- Check fixative expiration date prior to use. In doubt, use only fresh fixative.
4. Cause: Wrong fixative has been used.
Solution:
- 4% PFA is recommended.
- Make sure that the antibodies are specified to bind after methanol fixation. Keep in mind that some epitopes (like filament epitopes) are damaged by methanol.
5. Cause: Insufficient fixative volume, meaning that not enough active fixative was available or the tissue was not fully immersed.
Solution:
- Ensure you use a sufficiently large volume of fixative and ensure that the tissue is fully immersed.
6. Cause: Too short fixation time.
Solution:
- Increase the incubation time, but make sure to avoid over-fixation.
Organ degradation is commonly caused by under-fixation.
Under-fixation
1. Cause: Insufficient perfusion or missing post-fixation immersion perfusion.
Solution:
- If under-fixation is noticed early enough, an additional fixation step in the very beginning could help.
- No salvaging is possible after an organ is cleared; improve the fixation conditions for the next experiment.
2. Cause: Wrong fixative concentration
Solution: Use recommended concentration, e.g., 4% PFA.
3. Cause: Fixative is expired
Solution: Check fixative expiration date prior to use. In doubt, use only fresh fixative.
4. Cause: Wrong fixative has been used.
Solution:
- 4% PFA is recommended.
- Make sure that the antibodies are specified to bind after methanol fixation. Keep in mind that some epitopes (like filament epitopes) are damaged by methanol.
5. Cause: Insufficient fixative volume, meaning that not enough active fixative was available or the tissue was not fully immersed.
Solution: Ensure you use a sufficiently large volume of fixative and ensure that the tissue is fully immersed.
6. Cause: Too short fixation time.
Solution: Increase the incubation time, but make sure to avoid over-fixation.
Hazy autofluorescence can point to a common issue: under-fixation.
Under-fixation
1. Cause: Insufficient perfusion or missing post-fixation immersion perfusion.
Solution:
- If under-fixation is noticed early enough, an additional fixation step in the very beginning could help.
- No salvaging is possible after an organ is cleared; improve the fixation conditions for the next experiment.
2. Cause: Wrong fixative concentration
Solution: Use recommended concentration, e.g., 4% PFA.
3. Cause: Fixative is expired
Solution: Check fixative expiration date prior to use. In doubt, use only fresh fixative.
4. Cause: Wrong fixative has been used.
Solution:
- 4% PFA is recommended.
- Make sure that the antibodies are specified to bind after methanol fixation. Keep in mind that some epitopes (like filament epitopes) are damaged by methanol.
5. Cause: Insufficient fixative volume, meaning that not enough active fixative was available or the tissue was not fully immersed.
Solution: Ensure you use a sufficiently large volume of fixative and ensure that the tissue is fully immersed.
6. Cause: Too short fixation time.
Solution: Increase the incubation time, but make sure to avoid over-fixation.
When the visibility of anatomical structures or cellular positions becomes challenging or non-existent under a microscope, the underlying issue can be attributed to under-fixation.
Under-fixation
1. Cause: Insufficient perfusion or missing post-fixation immersion perfusion.
Solution:
- If under-fixation is noticed early enough, an additional fixation step in the very beginning could help.
- No salvaging is possible after an organ is cleared; improve the fixation conditions for the next experiment.
2. Cause: Wrong fixative concentration
Solution: Use recommended concentration, e.g., 4% PFA.
3. Cause: Fixative is expired
Solution: Check fixative expiration date prior to use. In doubt, use only fresh fixative.
4. Cause: Wrong fixative has been used.
Solution:
- 4% PFA is recommended.
- Make sure that the antibodies are specified to bind after methanol fixation. Keep in mind that some epitopes (like filament epitopes) are damaged by methanol.
5. Cause: Insufficient fixative volume, meaning that not enough active fixative was available or the tissue was not fully immersed.
Solution: Ensure you use a sufficiently large volume of fixative and ensure that the tissue is fully immersed.
6. Cause: Too short fixation time.
Solution: Increase the incubation time, but make sure to avoid over-fixation.
Tissue degradation and partial loss during or after fixation may be due to under-fixation.
Under-fixation
1. Cause: Insufficient perfusion or missing post-fixation immersion perfusion.
Solution:
- If under-fixation is noticed early enough, an additional fixation step in the very beginning could help.
- No salvaging is possible after an organ is cleared; improve the fixation conditions for the next experiment.
2. Cause: Wrong fixative concentration
Solution: Use recommended concentration, e.g., 4% PFA.
3. Cause: Fixative is expired
Solution: Check fixative expiration date prior to use. In doubt, use only fresh fixative.
4. Cause: Wrong fixative has been used.
Solution:
- 4% PFA is recommended.
- Make sure that the antibodies are specified to bind after methanol fixation. Keep in mind that some epitopes (like filament epitopes) are damaged by methanol.
5. Cause: Insufficient fixative volume, meaning that not enough active fixative was available or the tissue was not fully immersed.
Solution: Ensure you use a sufficiently large volume of fixative and ensure that the tissue is fully immersed.
6. Cause: Too short fixation time.
Solution: Increase the incubation time, but make sure to avoid over-fixation.
Organ shrinkage during sample preparation is usually due to over-fixation.
Over-fixation
If an organ is over-fixed, no salvaging is possible; improve the fixation conditions for the next experiment.
1. Cause: Too high fixative concentration or too long fixation time.
Solution: Decrease the fixative concentration or incubation time.
2. Cause: pH was not optimal.
Solution: Normally, for PFA the pH is adjusted to 7.2–7.4 and for the endogenous GFP preservation fixative to pH = 9. Dilute with HCl or NaOH for pH adjustment.
3. Cause: Temperature is too high.
Solution: Perform fixation at 4 °C and use cold fixative.
Non-specific antibody binding can have several causes. One reason, which occurs during the initial steps of sample preparation, is over-fixation. Of course, using validated and specific antibodies for a given marker is crucial as well. Check out our dedicated antibodies for 3D imaging of cleared tissues.
Over-fixation
If an organ is over-fixed, no salvaging is possible; improve the fixation conditions for the next experiment.
1. Cause: Too high fixative concentration or too long fixation time.
Solution: Decrease the fixative concentration or incubation time.
2. Cause: pH was not optimal.
Solution: Normally, for PFA the pH is adjusted to 7.2–7.4 and for the endogenous GFP preservation fixative to pH = 9. Dilute with HCl or NaOH for pH adjustment.
3. Cause: Temperature is too high.
Solution: Perform fixation at 4 °C and use cold fixative.
Exceptionally high autofluorescence is commonly caused by over-fixation.
Over-fixation
If an organ is over-fixed, no salvaging is possible; improve the fixation conditions for the next experiment.
1. Cause: Too high fixative concentration or too long fixation time.
Solution: Decrease the fixative concentration or incubation time.
2. Cause: pH was not optimal.
Solution: Normally, for PFA the pH is adjusted to 7.2–7.4 and for the endogenous GFP preservation fixative to pH = 9. Dilute with HCl or NaOH for pH adjustment.
3. Cause: Temperature is too high.
Solution: Perform fixation at 4 °C and use cold fixative.
The endogenous GFP signal is highly vulnerable, and its preservation heavily relies on the chosen tissue clearing method. However, other reasons such as over-fixation might also contribute to the loss of GFP fluorescence. Please also check our protocols if you are interested in the preservation of endogenous protein signals.
Over-fixation
If an organ is over-fixed, no salvaging is possible; improve the fixation conditions for the next experiment.
1. Cause: Too high fixative concentration / too long fixation time
Solution: Decrease the fixative concentration or incubation time
2. Cause: pH was not optimal
Solution: Normally for PFA, pH is adjusted to 7.2-7.4, for the endogenous GFP preservation fixative pH=9, dilute with HCL or NaOH for pH adjustment
3. Cause: Temperature is too high
Solution: Use 4°C fixative
Poor antibody penetration and labeling in comparison to other experiments using the same incubation time can be a result of over-fixation.
Over-fixation
If an organ is over-fixed, no salvaging is possible; improve the fixation conditions for the next experiment.
1. Cause: Too high fixative concentration or too long fixation time.
Solution: Decrease the fixative concentration or incubation time.
2. Cause: pH was not optimal.
Solution: Normally, for PFA the pH is adjusted to 7.2–7.4 and for the endogenous GFP preservation fixative to pH = 9. Dilute with HCl or NaOH for pH adjustment.
3. Cause: Temperature is too high.
Solution: Perform fixation at 4 °C and use cold fixative.
Dark residues observed during sample preparation are likely due to pigmented tissue or the presence of blood residues.
Tissue is pigmented or blood residues are present.
1. Cause: Perfusion was not done, or it was done inefficiently.
Solution: If possible, perform perfusion for the next experiment.
If perfusion is not possible, chemical bleaching might help to decolorize or depigment the area.
2. Cause: Too fast perfusion (can also introduce rupture artifacts).
Solution: Improve perfusion by utilizing a slower flow rate for the perfusion solution.
3. Cause: Perfusion solutions like heparin, PBS, EDTA are not able to efficiently remove the extra blood or other pigments which are not related to blood.
Solution: Try depigmentation / decolorization / chemical bleaching.
4. Cause: Perfusion is not possible due to age, or the tissue is already fixed.
Solution: Try depigmentation / decolorization / chemical bleaching.
5. Cause: Tissue-specific properties, i.e., pigmented tissues like skin, melanocytes, eyes.
Solution:
- Try depigmentation / decolorization / chemical bleaching.
- If possible, use a sample from a younger animal to have fewer pigments.
- Remove the pigmented tissue if it is not necessary.
6. Cause: Markers from medical operations, sample positioning marks, tumor cell tracers, contrast agents, tattoos pigments.
Solution: Physically remove markers if the area is not necessary for imaging.
If the sample falls out of the agarose or phytagel, agarose embedding failed.
Agarose embedding of the sample failed.
1. Cause: The surface of the organ was wet during embedding in agarose.
Solution: Dry the organ carefully with a small piece of tissue to avoid changes in concentration due to water content. Do not dry it out completely.
2. Cause: Temperature of the agarose or its concentration are not optimal.
Solution:
- Pay attention to the optimal temperature (as a rule of thumb, the bottle should be touchable by hand without hurting) and concentration of the agarose (1.5%) or phytagel (1%)
- Cover the organ completely if it is prone to falling out; re-embed it until it stays in.
Tissue rupture leading to imaging artifacts is the result of the tissue drying out.
Tissue is dried out.
1. Cause: Solution volumes are too low.
Solution: Make sure the sample is fully immersed in the respective liquids; store it in buffer after fixation.
2. Cause: Too fast dehydration.
Solution: No salvage is possible. Imaging can proceed, but may yield results with artifacts. Restart the process and ensure that the sample is floating, fully immersed, and does not stick to the tube. Slow down the rehydration process.
Tissue adhesion to the tube sidewalls during incubation is caused by the tissue drying out.
Tissue is dried out.
1. Cause: Solution volumes are too low.
Solution: Make sure the sample is fully immersed in the respective liquids; store it in buffer after fixation.
2. Cause: Too fast dehydration.
Solution: No salvage is possible. Imaging can proceed, but may yield results with artifacts. Restart the process and ensure that the sample is floating, fully immersed, and does not stick to the tube. Slow down the rehydration process.
Pronounced autofluorescence signals can be due to dried-out tissue or over-fixation.
Tissue is dried out.
1. Cause: Solution volumes are too low.
Solution: Make sure the sample is fully immersed in the respective liquids; store it in buffer after fixation.
2. Cause: Too fast dehydration.
Solution: No salvage is possible. Imaging can proceed, but may yield results with artifacts. Restart the process and ensure that the sample is floating, fully immersed, and does not stick to the tube. Slow down the rehydration process.
Over-fixation
If an organ is over-fixed, no salvaging is possible; improve the fixation conditions for the next experiment.
1. Cause: Too high fixative concentration / too long fixation time.
Solution: Decrease the fixative concentration or incubation time.
2. Cause: pH was not optimal.
Solution: Normally for PFA, pH is adjusted to 7.2-7.4, for the endogenous GFP preservation fixative pH=9, dilute with HCL or NaOH for pH adjustment.
3. Cause: Temperature is too high.
Solution: Use 4°C fixative.
Labeling
Strong staining on the surface of tissue / staining gradient
1. Cause: Insufficient antibody incubation.
Solution: Increase the incubation time according to the tissue size.
2. Cause: Insufficient permeabilization.
Solution:
- Increase time for incubation with permeabilization solution (sample specific).
- Refresh permeabilization solution, and increase solution volume according to the tissue size.
- Improve permeabilization by an additional dehydration/rehydration series using 2% Tween®-20 in ethanol prior to staining.
3. Cause: Insufficient penetration or diffusion of secondary antibody.
Solution:
- Use conjugated primary antibodies to reach a homogeneous labeling and penetration of the sample and to reduce the overall protocol time in parallel.
- Increase incubation time for the secondary antibody or try out a different secondary antibody.
- Optimize the permeabilization (see ''Insufficient permeabilization'').
4. Cause: High antibody concentration.
Solution:
- Optimize antibody concentration via titration and consider lower concentrations.
- For Miltenyi Biotec antibodies, refer to their recommended dilution.
- For other providers include dilution ranges given for IHC.
5. Cause: Insufficient antibody amount; depletion of antibody by target.
Solution:
- Optimize antibody concentration via titration and consider higher concentrations.
- For Miltenyi Biotec antibodies, refer to their recommended dilutions.
- For other providers include dilution ranges given for IHC.
6. Cause: Insufficient blocking.
Solution:
- Consider a longer blocking incubation.
- Increase blocking serum volume.
7. Cause: Antibody aggregates on sample surface.
Solution:
- If the aggregates are present in the antibody stock solution, centrifugation (e.g. 10,000×g for 10 min) or filtration might help.
- If penetration is not causing this issue, reduce the antibody concentration
8. Cause: Antibody is too large or unsuited and cannot penetrate the sample properly.
Solution: Try other antibodies if available.
9. Cause: A natural matter of high abundance of antigens on the surface of the tissue, for example, abundance of astrocytes and their markers in comparison to other glial cells or neurons.
10. Cause: The antibody does not stain the region of interest but other regions on the surface.
Solution:
- Increase the antibody incubation time.
- Increase antibody concentration and optimize permeabilization (see ''Insufficient permeabilization'').
Over-fixation
If organ is over-fixed, no salvaging is possible; improve for the next experiment
1. Cause: Too high fixative concentration / too long fixation time
Solution: Decrease the fixative concentration or incubation time
2. Cause: pH was not optimal
Solution: Normally for PFA, pH is adjusted to 7.2-7.4, for the endogenous GFP preservation fixative pH=9, dilute with HCL or NaOH for pH adjustment
3. Cause: Temperature is too high
Solution: Use 4°C fixative
An uneven signal pattern may be due to strong staining on the tissue surface, the presence of a staining gradient, or over-fixation.
Strong staining on the surface of tissue / staining gradient
1. Cause: Insufficient antibody incubation.
Solution: Increase the incubation time according to the tissue size.
2. Cause: Insufficient permeabilization.
Solution:
- Increase time for incubation with permeabilization solution (sample specific).
- Refresh permeabilization solution, and increase solution volume according to the tissue size.
- Improve permeabilization by an additional dehydration/rehydration series using 2% Tween®-20 in ethanol prior to staining.
3. Cause: Insufficient penetration or diffusion of secondary antibody.
Solution:
- Use conjugated primary antibodies to reach a homogeneous labeling and penetration of the sample and to reduce the overall protocol time in parallel.
- Increase incubation time for the secondary antibody or try out a different secondary antibody.
- Optimize the permeabilization (see ''Insufficient permeabilization'').
4. Cause: High antibody concentration.
Solution:
- Optimize antibody concentration via titration and consider lower concentrations.
- For Miltenyi Biotec antibodies, refer to their recommended dilution.
- For other providers include dilution ranges given for IHC.
5. Cause: Insufficient antibody amount; depletion of antibody by target.
Solution:
- Optimize antibody concentration via titration and consider higher concentrations.
- For Miltenyi Biotec antibodies, refer to their recommended dilutions.
- For other providers include dilution ranges given for IHC.
6. Cause: Insufficient blocking.
Solution:
- Consider a longer blocking incubation.
- Increase blocking serum volume.
7. Cause: Antibody aggregates on sample surface.
Solution:
- If the aggregates are present in the antibody stock solution, centrifugation (e.g. 10,000×g for 10 min) or filtration might help.
- If penetration is not causing this issue, reduce the antibody concentration
8. Cause: Antibody is too large or unsuited and cannot penetrate the sample properly.
Solution: Try other antibodies if available.
9. Cause: A natural matter of high abundance of antigens on the surface of the tissue, for example, abundance of astrocytes and their markers in comparison to other glial cells or neurons.
10. Cause: The antibody does not stain the region of interest but other regions on the surface.
Solution:
- Increase the antibody incubation time.
- Increase antibody concentration and optimize permeabilization (see ''Insufficient permeabilization'').
Over-fixation
If organ is over-fixed, no salvaging is possible; improve for the next experiment
1. Cause: Too high fixative concentration / too long fixation time
Solution: Decrease the fixative concentration or incubation time
2. Cause: pH was not optimal
Solution: Normally for PFA, pH is adjusted to 7.2-7.4, for the endogenous GFP preservation fixative pH=9, dilute with HCL or NaOH for pH adjustment
3. Cause: Temperature is too high
Solution: Use 4°C fixative
The absence of a visible signal in imaging can be the result of either weak or no staining, or under-fixation.
Weak staining or no staining
1. Cause: Wrong fixative.
Solution: Use 4% PFA.
2. Cause: Antibody is not suitable.
Solution: Try other antibodies if available.
3. Cause: Degradation of the epitopes during fixation, depigmentation (decolorization, chemical bleaching), or clearing.
Solution: Tissue might be old; under-fixation happened; see "Under-fixation" section.
4. Cause: H₂O₂ depigmentation can affect epitopes (for example mouse CD31).
Solution: Other clearing methods might be suitable.
5. Cause: Light exposure, photobleaching.
Solution: Avoid light exposure throughout sample preparation process.
Consider using Miltenyi Biotec’s 3D-IF antibodies conjugated to highly photostable Vio® Dyes.
6. Cause: Marker is not expressed on the sample.
Solution: Check if the absence of staining or marker has a biological relevance or is due to technical issues. Include a positive control for tissue staining in parallel to your staining.
7. Cause Antibody/reagents are too old.
Solution: Check the expiry date of the product before use.
8. Cause: Primary antibody labeling resulted in a weak signal.
Solution: To amplify the signal, different staining strategies can be pursued: (a) Using a primary non-conjugated antibody + secondary antibody; (b) using a polyclonal antibody to bind more epitopes per target protein; (c) increasing the antibody concentration if non-specific binding is not an issue.
9. Cause: Quenching of GFP fluorescence or other fluorescent proteins.
Solution: Adapt the protocol by adjusting dehydration agent and pH to preserve the GFP signal, or use anti-GFP antibodies to boost the signal.
Use the protocol from Miltenyi Biotec to preserve endogenous fluorescent protein signal in mouse brain hemisphere.
Under-fixation
1. Cause: Insufficient perfusion or missing post-fixation immersion perfusion.
Solution:
- If under-fixation is noticed early enough, an additional fixation step in the very beginning could help.
- No salvaging is possible after organ is cleared; improve the fixation conditions for the next experiment.
2. Cause: Wrong fixative concentration
Solution: Use recommended concentration, e.g., 4% PFA.
3. Cause: Fixative is expired.
Solution: Check fixative expiration date prior to use. If in doubt, use only fresh fixative.
4. Cause: Wrong fixative has been used.
Solution:
- 4% PFA is recommended.
- Make sure that the antibodies are specified to bind after methanol fixation. Keep in mind that some epitopes (like filament epitopes) are damaged by methanol,
5. Cause: Insufficient fixative volume, meaning that not enough active fixative was available or the tissue was not fully immersed.
Solution: Use a large enough volume of fixative and make sure that the tissue is fully immersed.
6. Cause: Too short fixation time.
Solution: Increase the incubation time, but make sure to avoid over-fixation.
If only an autofluorescence signal is visible, this may be due to weak staining, absence of staining altogether or under-fixation.
Weak staining or no staining
1. Cause: Wrong fixative.
Solution: Use 4% PFA.
2. Cause: Antibody is not suitable.
Solution: Try other antibodies if available.
3. Cause: Degradation of the epitopes during fixation, depigmentation (decolorization, chemical bleaching), or clearing.
Solution: Tissue might be old; under-fixation happened; see "Under-fixation" section.
4. Cause: H₂O₂ depigmentation can affect epitopes (for example mouse CD31).
Solution: Other clearing methods might be suitable.
5. Cause: Light exposure, photobleaching.
Solution: Avoid light exposure throughout sample preparation process.
Consider using Miltenyi Biotec’s 3D-IF antibodies conjugated to highly photostable Vio® Dyes.
6. Cause: Marker is not expressed on the sample.
Solution: Check if the absence of staining or marker has a biological relevance or is due to technical issues. Include a positive control for tissue staining in parallel to your staining.
7. Cause Antibody/reagents are too old.
Solution: Check the expiry date of the product before use.
8. Cause: Primary antibody labeling resulted in a weak signal.
Solution: To amplify the signal, different staining strategies can be pursued: (a) Using a primary non-conjugated antibody + secondary antibody; (b) using a polyclonal antibody to bind more epitopes per target protein; (c) increasing the antibody concentration if non-specific binding is not an issue.
9. Cause: Quenching of GFP fluorescence or other fluorescent proteins.
Solution: Adapt the protocol by adjusting dehydration agent and pH to preserve the GFP signal, or use anti-GFP antibodies to boost the signal.
Use the protocol from Miltenyi Biotec to preserve endogenous fluorescent protein signal in mouse brain hemisphere.
Under-fixation
1. Cause: Insufficient perfusion or missing post-fixation immersion perfusion.
Solution:
- If under-fixation is noticed early enough, an additional fixation step in the very beginning could help.
- No salvaging is possible after organ is cleared; improve the fixation conditions for the next experiment.
2. Cause: Wrong fixative concentration
Solution: Use recommended concentration, e.g., 4% PFA.
3. Cause: Fixative is expired.
Solution: Check fixative expiration date prior to use. If in doubt, use only fresh fixative.
4. Cause: Wrong fixative has been used.
Solution:
- 4% PFA is recommended.
- Make sure that the antibodies are specified to bind after methanol fixation. Keep in mind that some epitopes (like filament epitopes) are damaged by methanol,
5. Cause: Insufficient fixative volume, meaning that not enough active fixative was available or the tissue was not fully immersed.
Solution: Use a large enough volume of fixative and make sure that the tissue is fully immersed.
6. Cause: Too short fixation time.
Solution: Increase the incubation time, but make sure to avoid over-fixation.
High background signals can have various causes, including over-fixation and dried-out tissue, for example.
Over-fixation
If organ is over-fixed, no salvaging is possible; improve for the next experiment
1. Cause: Too high fixative concentration / too long fixation time
Solution: Decrease the fixative concentration or incubation time
2. Cause: pH was not optimal
Solution: Normally for PFA, pH is adjusted to 7.2-7.4, for the endogenous GFP preservation fixative pH=9, dilute with HCL or NaOH for pH adjustment
3. Cause: Temperature is too high
Solution: Use 4°C fixative
Tissue is dried out
1. Cause: Solution volumes are too low.
Solution: Make sure the sample is fully immersed in the respective liquids; store it in buffer after fixation.
2. Cause: Too fast dehydration.
Solution: No salvaging possible. Imaging can be done but might yield results with artifacts. Restart and make sure that the sample is floating, is fully immersed, and does not stick to the tube. Slow down the rehydration.
Further causes of high background signal
1. Cause: Insufficient washing steps.
Solution: Add more or longer washing steps.
2. Cause: Antibody concentration is too high.
Solution: Use the recommended concentration; titrate antibody down.
3. Cause: Tissue-specific issues related to non-specific antibody binding or stickiness. Solution:
- Test non-specific antibody binding using an isotype control antibody.
- If antibody penetration is not the cause of the issue, non-specific binding might be a feature of this particular antibody/sample combination.
4. Cause: Insufficient blocking.
Solution:
- Optimize blocking see Strong staining on the surface of tissue or staining gradient: Insufficient blocking.
- Change antibody if it still produces background.
5. Cause: Autofluorescence.
Solution:
- Consider depigmentation (decolorization, chemical bleaching) protocols which might reduce autofluorescence (e.g. melanin in skin).
- Sample the tissue from younger animals which tend to have less autofluorescence-producing pigments.
- Physically remove the autofluorescent sample parts, e.g., skin.
Scattered fluorescent signals on the tissue may have several causes such as antibody aggregates, air bubbles, contamination, etc.
Scattered fluorescent signal on the tissue
1. Cause: Aggregates from the antibody solution.
Solution: If the aggregates are present in the antibody stock solution, centrifugation (e.g. 10,000×g for 10 min) or filtration might help.
2. Cause: Buffer contamination.
Solution: Filter the buffer or simply use fresh buffers.
3. Cause: Air bubbles; not a real fluorescent signal, rather ambient light, is reflected in all channels.
Solution:
- Gently shake the tissue in its solution.
- Remove the bubbles using forceps or a desiccator (vacuum extraction of air bubbles).
- Extract air bubbles, e.g., from lungs using insulin syringe by suction or by injection of liquid.
4. Cause: Necrotic areas.
Solution:
- Choose bright fluorochromes that exceed the background signal.
- Exclude this area from imaging.
- Cut away necrotic regions.
- If possible, exclude the area from the image analysis.
Tissue clearing
Blurry tissue images or autofluorescence signals can result from incomplete tissue clearing, over-fixation, or pigmentation / blood residue.
Tissue is not cleared.
1. Cause: Insufficient permeabilization.
Solution: Optimize the permeabilization (see Strong staining on the surface of tissue or staining gradient: Insufficient permeabilization).
2. Cause: Insufficient tissue clearing or dehydration solution volume.
Solution: Make sure the solution volume is large enough to cover the tissue and is also sufficient for the chemical dehydration / clearing processes.
3. Cause: Insufficient incubation time.
Solution: Increase the incubation time relative to the sample size; see Miltenyi Biotec’s tissue clearing protocols.
4. Cause: Suboptimal washing steps.
Solution: Always make sure you include long enough washing steps with sufficient volume of washing buffers.
5. Cause: Refractive index (RI) is altered.
Solution:
- Make sure not to contaminate the clearing solution with other substances. If in doubt, use fresh clearing solution.
- The sample might carry a bit of ethanol after dehydration. If the clearing solution volume is low, then the ethanol mixing with clearing solution might lead to RI mismatch. Make sure you use an appropriate volume of clearing solution.
6. Cause: Clearing reagents are expired.
Solution: Check the shelf life.
7. Cause: Temperature is too low.
Solution: Increase temperature to about 28 °C, if possible. Verify if antibody staining resists higher temperature during clearing step.
8. Cause: Dehydration was too short.
Solution:
If dehydration was too short and minimal residual water remains in the sample, it becomes blurry (not fully white). A second round of incubation in clearing solution can be helpful to render the sample transparent. Also see Tissue turns white during clearing.
- Increase dehydration time and/or volume.
9. Cause: Insufficient delipidation.
Solution:
- Optimize the permeabilization (see Strong staining on the surface of tissue or staining gradient: Insufficient permeabilization).
- Increase the incubation time with ethanol series.
- Increase solution volume according to the tissue size.
Tissue is pigmented or blood residues are present.
1. Cause: Perfusion was not done, or it was done inefficiently.
Solution: If possible, perform perfusion for the next experiment.
If perfusion is not possible, chemical bleaching might help to decolorize or depigment the area.
2. Cause: Too fast perfusion (can also introduce rupture artifacts)
Solution: Improve perfusion by using a slower perfusion solution flow.
3. Cause: Perfusion solutions like heparin, PBS, EDTA are not able to efficiently remove the extra blood or other pigments which are not related to blood.
Solution: Try depigmentation / decolorization / chemical bleaching.
4. Cause: Perfusion is not possible due to age, or the tissue is already fixed.
Solution: Try depigmentation / decolorization / chemical bleaching.
5. Cause: Tissue-specific properties, i.e., pigmented tissues like skin, melanocytes, eyes.
Solution:
- Try depigmentation / decolorization / chemical bleaching.
- If possible, use a sample from a younger animal to have fewer pigments.
- Remove the pigmented tissue if it is not necessary.
6. Cause: Markers from medical operations, sample positioning marks, tumor cell tracers, contrast agents, tattoos pigments.
Solution: Physically remove markers if the area is not necessary for imaging.
Over-fixation
If organ is over-fixed, no salvaging is possible; improve for the next experiment
1. Cause: Too high fixative concentration / too long fixation time
Solution: Decrease the fixative concentration or incubation time
2. Cause: pH was not optimal
Solution: Normally for PFA, pH is adjusted to 7.2-7.4, for the endogenous GFP preservation fixative pH=9, dilute with HCL or NaOH for pH adjustment
3. Cause: Temperature is too high
Solution: Use 4°C fixative
Foggy samples can be the result of insufficient sample clearing, over-fixation, or pigmentation/blood residues.
Tissue is not cleared.
1. Cause: Insufficient permeabilization.
Solution: Optimize the permeabilization (see Strong staining on the surface of tissue or staining gradient: Insufficient permeabilization'').
2. Cause: Insufficient tissue clearing or dehydration solution volume.
Solution: Make sure the solution volume is large enough to cover the tissue and is also sufficient for the chemical dehydration / clearing processes.
3. Cause: Insufficient incubation time.
Solution: Increase the incubation time relative to the sample size; see Miltenyi Biotec’s tissue clearing protocols.
4. Cause: Suboptimal washing steps.
Solution: Always make sure you include long enough washing steps with sufficient volume of washing buffers.
5. Cause: Refractive index (RI) is altered.
Solution:
- Make sure not to contaminate the clearing solution with other substances. If in doubt, use fresh clearing solution.
- The sample might carry a bit of ethanol after dehydration. If the clearing solution volume is low, then the ethanol mixing with clearing solution might lead to RI mismatch. Make sure you use an appropriate volume of clearing solution.
6. Cause: Clearing reagents are expired.
Solution: Check the shelf life.
7. Cause: Temperature is too low.
Solution: Increase temperature to about 28 °C, if possible. Verify if antibody staining resists higher temperature during clearing step.
8. Cause: Dehydration was too short.
Solution:
If dehydration was too short and minimal residual water remains in the sample, it becomes blurry (not fully white). A second round of incubation in clearing solution can be helpful to render the sample transparent. Also see Tissue turns white during clearing.
- Increase dehydration time and/or volume.
9. Cause: Insufficient delipidation.
Solution:
- Optimize the permeabilization (see Strong staining on the surface of tissue or staining gradient: Insufficient permeabilization).
- Increase the incubation time with ethanol series.
- Increase solution volume according to the tissue size.
Tissue is pigmented or blood residues are present.
1. Cause: Perfusion was not done, or it was done inefficiently.
Solution: If possible, perform perfusion for the next experiment.
If perfusion is not possible, chemical bleaching might help to decolorize or depigment the area.
2. Cause: Too fast perfusion (can also introduce rupture artifacts)
Solution: Improve perfusion by using a slower perfusion solution flow.
3. Cause: Perfusion solutions like heparin, PBS, EDTA are not able to efficiently remove the extra blood or other pigments which are not related to blood.
Solution: Try depigmentation / decolorization / chemical bleaching.
4. Cause: Perfusion is not possible due to age, or the tissue is already fixed.
Solution: Try depigmentation / decolorization / chemical bleaching.
5. Cause: Tissue-specific properties, i.e., pigmented tissues like skin, melanocytes, eyes.
Solution:
- Try depigmentation / decolorization / chemical bleaching.
- If possible, use a sample from a younger animal to have fewer pigments.
- Remove the pigmented tissue if it is not necessary.
6. Cause: Markers from medical operations, sample positioning marks, tumor cell tracers, contrast agents, tattoos pigments.
Solution: Physically remove markers if the area is not necessary for imaging.
Over-fixation
If organ is over-fixed, no salvaging is possible; improve for the next experiment
1. Cause: Too high fixative concentration / too long fixation time
Solution: Decrease the fixative concentration or incubation time
2. Cause: pH was not optimal
Solution: Normally for PFA, pH is adjusted to 7.2-7.4, for the endogenous GFP preservation fixative pH=9, dilute with HCL or NaOH for pH adjustment
3. Cause: Temperature is too high
Solution: Use 4°C fixative
The tissue is probably not sufficiently dehydrated.
Tissue turns white during clearing.
1. Cause: The tissue is not sufficiently dehydrated.
Solution:
- Not salvageable if sample contains substantial amounts of residual water; improve parameters as indicated below for the next experiments.
- If dehydration was too short and minimal residual water remains in sample, it becomes blurry (not fully white). A second round of incubation in clearing solution can be helpful to render the sample transparent.
- Consider adding multiple or longer dehydration steps.
- Volumes of dehydration or clearing solution are insufficient for the size of the sample.
- Increase the dehydration duration.
- If the organ has floated up during dehydration or clearing (they do after transfer from 100% ethanol to clearing solution), gently shake the sample after transfer to the clearing solution. Make sure it can sink down; if it doesn't, one part of the organ turns white.
- Make sure that the sample does not create streaks in the ethanol series or clearing solution anymore before imaging or transfer to the next tube. If streaks persist, repeat the current step and exchange for fresh ethanol or clearing solution.
2. Cause: Ethanol contains water.
Solution:
- Use fresh ethanol bottle.
- We recommend adding activated/dehydrated molecular sieves (3 Å, 4–8 mesh, Sigma, 208574-1-KG) to stored 2.5 L bottles of pure ethanol.
Imaging
Blurry spots on the sample can have several causes, such as a mismatch of the refractive index (RI) or insufficient delipidation.
Blurry spots
1. Cause: RI mismatch or insufficient delipidation.
Solution: Follow the source of blurriness. Based on the tissue type and your observations you may find a relevant cause and solution on this webpage.
2. Cause: Burned tissue from cauterization.
Solution:
- Try not to burn the tissue.
- Cut away the burned part. The burned tissue is not salvageable.
3. Cause: Air bubbles stuck in the sample.
Solution: The emission light can be deflected and lost, see Scattered fluorescent signal on the tissue section above.
You have previously chosen International as your location.