The BCR RNA Profiling Kit, human supports an RNA input equivalent to 100 to 500,000 sorted or purified memory/naive B cells, 10 to 1,000 plasma cells/plasmablasts, from Peripheral blood mononuclear cells (PBMCs), or from B cell–containing tissues, and produces indexed, ready-to-sequence libraries for Illumina®‑based sequencing. For sorted PBMCs and B Cells, this protocol also works directly from cells, allowing users to skip prior RNA isolation (Figure 1).
Generation of IG heavy- and light-chain cDNA libraries using the BCR RNA Profiling Kit, human
This application protocol describes the generation of IG heavy- and light‑chain cDNA libraries using a combination of highly sensitive multiplex PCR and Unique Molecular Identifiers (UMIs) with the BCR RNA Profiling Kit, human. UMIs are introduced along with first‑strand cDNA synthesis and enable error‑free reconstruction of full‑length Immunoglobulin (IG) heavy- and light‑chain sequences with hypermutations (except in the FR1 region). In addition, UMIs enable for quantification of template cDNA molecules, control of input bottlenecks, and accurate normalization of samples for comparative analysis. Gene‑specific primers targeting IG heavy‑chain constant regions facilitate discrimination of all IGH isotypes, including IgM, IgD, IgG3, IgG1, IgA1, IgG2, IgG4, IgE, and IgA2. For details regarding the required materials and instruments, please go to the Materials tab.
General recommendations to prevent contamination
General recommendations, such as using lab coats, gloves, tips with aerosol filters, certified RNAse/DNAse-free reagents, and decontamination of the workspacebefore working with RNA (e.g. using AppliChem RNase-ExitusPlus™), to minimize the risk of RNA degradation and contamination should be followed.
Performing non-template control reactions is strongly recommended for each experiment. To prevent cross-sample contamination, library preparation should be performed in two physically separated workspaces (PCR boxes): one for pre–PCR and one for post–PCR, located in different rooms. Avoid transferring anything (reagents, pipettes, racks, etc.) from the post–PCR to the pre–PCR workspace. In the case of possible contamination, use decontaminating procedures with special reagents (e.g. using AppliChem DNA/RNA-ExitusPlus).
Cell isolation, RNA purification, cDNA synthesis, first- and second-round PCR master mix preparations should be performed in the pre–PCR workspace. After the first PCR amplification, the tubes should be transferred to the post–PCR workspace to perform all remaining steps except PCR master mix preparation.
Starting material
As starting material, isolated RNA from sorted/isolated B cells, PBMCs, or B cell–containing tissues can be used. Furthermore, it is possible to start directly from intact PBMCs or B cells.
When starting directly from cells, no RNA isolation is needed. Cells should be pelleted by centrifugation and kept on ice during processing. Direct-from-cells procedure was tested for fresh and frozen samples containing up to 20,000 PBMCs.
When starting with RNA, it is necessary to use a Qubit® fluorimeter or other fluorescent-based methods to determine the RNA concentration. High RNA quality is critical for efficient library preparation. Residual traces of proteins, salts, or other contaminants may decrease the efficiency of the enzymatic activity necessary for optimal targeted enrichment. Refer to section "5. Appendix A: Preparing starting material" for more details about recommendations regarding RNA quality for use as input to the protocol. gDNA can significantly affect the cDNA synthesis reaction and subsequent PCR amplification. It is strongly recommended to perform DNase treatment during RNA extraction.
For quantitative IG repertoire profiling, we recommend separating B cell populations with different IG RNA transcription levels (i.e. plasma cells/plasmablasts and memory/naive B cells) by magnetic-activated cell sorting.
In appendix A provides recommendations for RNA Isolation under the following conditions:
- RNA isolation for large samples (>50,000 cells)
- RNA isolation for small samples (i.e. <50,000 cells)
If further guidance is needed, some frequently occurring situations are described in appendix A and help to choose the appropriate strategy for RNA isolation from:
- PBMCs
- 100 to 50,000 sorted or purified B cells
- more than 50,000 sorted or purified B cells
- B cell–containing tissues
It is also possible to use TRIzolTM reagent to isolate RNA.
Please refer to appendix B for a detailed protocol.
Additional information
Before starting the experiment, it is recommended to read the datasheet of the BCR RNA Profiling Kit, human. Also read the specifications of the necessary materials and instruments that are described on the Materials tab of the protocol webpage.
Start of the experiment
- Perform cDNA synthesis, clean-up after cDNA synthesis, and the PCR1 and PCR2 master mix preparation in a pre–PCR workspace.
- All samples and reagents from the BCR RNA Profiling Kit Box 1 must be kept on ice during reaction set-up. After use, immediately return enzymes to the freezer.
- Clean-up steps are performed at room temperature (+15 to +25 °C).
- Thoroughly mix all components before use.
- For mixing by pipetting, especially at low volumes, it is advised to set the pipette volume to approximately 80% to avoid creating bubbles.
cDNA synthesis starting from RNA
Before start: Thaw RNA samples on ice. Thaw the BCR cDNA Synthesis Primers briefly at room temperature (+15 to +25 °C), mix primers by vortexing, and spin down. Take out the RT Master Mix from the –20 °C freezer, mix briefly by vortexing, and spin down. Place all components on ice until use.
1. In nuclease-free PCR tubes, mix the following components in a final reaction volume of 12 μL per reaction according to table 1.
Table 1: cDNA Synthesis Primer Mix
| Component | Volume (in μL) |
| RNA (0.1–500 ng per reaction) | 1–10 |
| BCR cDNA Synthesis Primers | 2 |
| Water | 0–9 |
| Total volume per reaction | 12 |
2. Pre-heat a thermal cycler at +70 °C. Incubate the samples for 2 minutes at +70 °C, followed by 2 minutes at +4 °C.
3. During incubation, prepare the Reverse Transcription mix in a final volume of 8 µL per reaction according to table 2.
Table 2: Reverse Transcription mix
| Component | Volume (in μL) |
| RT Master Mix | 4 |
| Water | 4 |
| Total volume per reaction | 8 |
4. After incubation, add 8 μL of the Reverse Transcription mix to each sample, mix well by pipetting up and down 15 times without creating bubbles. Spin briefly and incubate for 30 minutes at +55 °C (lid temperature set to +60 °C).
5. Proceed with clean-up after cDNA synthesis (see chapter Clean-up after cDNA synthesis).
cDNA synthesis starting directly from cells
The direct‑from‑cells procedure was tested on fresh and frozen samples containing up to 20,000 PBMCs.
Before start: Thaw the BCR cDNA Synthesis Primers briefly at room temperature (+15 to +25 °C). Mix each component thoroughly by pipetting up and down before use, and spin down. Take the RT Master Mix from –20 °C freezer, mix briefly by vortexing, and spin down. Place all components on ice until use.
For working with input of under 1,000 B cells, use cells that have been sorted directly into a tube containing cDNA Synthesis master mix. Skip the pelleting step to avoid the risk of losing the sample and start with step 7. For frozen samples (cell pellets), also start with step 7.
1. Transfer cells into a nuclease-free 1.5 mL tube.
2. Centrifuge at 350 × g for 5 minutes at room temperature (+15 to +25 °C).
3. Wash the cells once with 1 mL PBS without Ca2+, Mg2+, or BSA and repeat centrifugation at 350 × g for 5 minutes.
4. Remove the supernatant by carefully removing 900 µL using a 1 mL pipette tip.
5. Continue to remove the supernatant by carefully removing approximately 90 µL using a 100 or 200 µL pipette tip. Be careful not to touch the side of the tube where the pellet is located.
▲ Note: The remaining PBS will not influence the reaction performance.
6. Place the tube on ice.
7. Prepare the cDNA Synthesis master mix for cells of the following components in a final volume of 16 µL per reaction according to table 3.
Table 3: cDNA Synthesis master mix for cells
| Component | Volume (in μL) |
| RT Master Mix | 4 |
| Water | 5 |
| Lysis Solution | 5 |
| BCR cDNA Synthesis Primers | 2 |
| Total volume per reaction | 16 |
8. If starting from frozen cell pellets, thaw the samples on ice until no frozen sample material is visible.
▲ Note: If starting with cells sorted directly into the cDNA Synthesis master mix, sort the target cell population straight into the tube containing the cDNA Synthesis master mix. Up to 1,000 cells can be sorted into the master mix volume of 16 µL, provided the total droplet volume does not exceed 4 µL. After sorting, briefly spin down the tube to collect any droplets from the walls of the tube. Then gently pipette up and down 10 times, avoiding bubble formation, and proceed to step 10.
9. Pipette 16 µL of the cDNA Synthesis master mix to the tube containing cell pellet. Mix carefully by pipetting up and down 10 times without creating bubbles.
10. Transfer the samples into the corresponding tube of a PCR strip, keep on ice until all samples are transferred.
11. Centrifuge briefly and incubate the samples for 30 minutes at +55 °C (lid temperature set to +60 °C).
12. Proceed with clean-up after cDNA synthesis
Clean-up after cDNA synthesis
Purify the obtained cDNA synthesis products using AMPure® XP Beads or SPRIselectTM reagent (Beckman CoulterTM) as follows:
Before start: Freshly prepare 80% ethanol and adjust beads to room temperature according to manufacturer’s instructions.
- Before use, thoroughly vortex the beads until homogenized and no pellet remains at the bottom of the bottle.
- Add 20 μL of beads to each reaction and mix well by pipetting up and down 10 times without creating bubbles.
- Incubate the solution for 5 minutes at room temperature (+15 to +25 °C).
- Place the tubes on the magnetic rack and wait until the solution becomes clear blue.
- While the tubes are on the magnetic rack, carefully remove the supernatant.
- Keep the tubes on the magnetic rack and add 200 μL of freshly prepared 80% ethanol. Wait for 30 seconds and remove the ethanol.
- Remove tubes from the magnetic rack. Briefly spin the tubes down and place them back on the magnetic rack. Remove all the residual ethanol carefully, using a small pipette.
- Remove tubes from the magnetic rack. Add 21 μL of water to the beads, mix well by pipetting up and down 15 times without creating bubbles, and incubate for 5 minutes at room temperature (+15 to +25 °C).
- Place the tubes on the magnetic rack and wait for the solution to clear.
- While the tubes are on the magnetic rack, carefully transfer 20 µL of the supernatant into a new tube.
- Homogenize beads by vortexing thoroughly, as described in step 1.
- Add 20 μL beads and mix well by pipetting up and down 10 times without creating bubbles.
- Incubate the solution for 5 minutes at room temperature (+15 to +25 °C).
- Place the tubes on the magnetic rack and wait for the solution to clear.
- While tubes are on the magnetic rack, carefully remove and discard the supernatant.
- Add 200 μL of freshly prepared 80% ethanol to the tubes, while they remain on the magnetic rack. Wait for 30 seconds and remove the ethanol.
- Remove tubes from the magnetic rack. Briefly spin the tubes down and place them back on the magnetic rack. Remove all the residual ethanol carefully, using a small pipette.
▲ Note: It is important to remove all traces of ethanol at this step as it could inhibit subsequent PCR reactions. - Remove tubes from the magnetic rack. Add 21 μL of Elution Buffer to the beads for elution and mix well by pipetting up and down without creating bubbles.
- Incubate the solution for 5 minutes at room temperature (+15 to +25 °C).
- Place tubes on the magnetic rack and wait for the solution to clear.
- Keep tubes on the magnetic rack and carefully transfer 20 μL of the cDNA into a new nuclease-free reaction tube.
▲ Note: During the elution steps, beads clumping can significantly reduce the cDNA yield. In case bead clumping is observed, please refer to the chapter Troubleshooting - Decreased cDNA synthesis / yield.
First PCR amplification
Before start: Thaw BCR-H PCR1 and BCR-K/L PCR1 Primers briefly at room temperature (+15 to +25 °C) and PCR Master Mix on ice. Mix each component thoroughly by pipetting up and down before use and spin down. Place all components on ice until use.
▲ Note: To adjust for the 2−3-fold higher BCR-K/L mRNA expression levels, it is recommended to use 15 µL eluate for BCR-H and 5 µL for BCR-K/L reaction as input. All further protocol steps are based on this experimental setup.
▲ Note: Depending on the scope of the experiment, the cDNA input for PCR1 may be adjusted between 0 and 20 µL of cDNA eluate for the BCR‑H and BCR‑K/L reactions, respectively. For example, if only BCR‑H is of interest, the entire 20 µL of cDNA eluate may be used for the BCR‑H PCR1, skipping the BCR‑K/L reaction. Please ensure that the volume of water is adjusted accordingly. However, modifying the cDNA input may require additional optimization of PCR cycle numbers and adaptation of sequencing recommendations.
1. At this point, each cDNA sample must be split into two nuclease-free tubes: 15 µL cDNA for BCR-H in one tube and 5 µL cDNA for BCR-K/L library in another tube.
2. For the BCR-H PCR1 master mix, use a final volume of 35 µL per reaction:
Table 4: BCR-H PCR1 master mix
| Component | Volume (in µL) |
| Water | 8 |
| PCR Master Mix | 25 |
| BCR-H PCR1 Primers | 2 |
| Total volume per reaction | 35 |
3. For the BCR-K/L PCR1 master mix, use a final volume of 45 µL per reaction.
Table 5: BCR-K/L PCR1 master mix
| Component | Volume (in µL) |
| Water | 18 |
| PCR Master Mix | 25 |
| BCR-K/L PCR1 Primers | 2 |
| Total volume per reaction | 45 |
4. Add the appropriate master mix to the purified cDNA samples:
- Add 35 μL of BCR-H PCR1 master mix to 15 μL cDNA sample. Mix gently by pipetting 10 times without creating bubbles and centrifuge briefly.
- Add 45 μL of BCR-K/L PCR1 master mix to 5 μL cDNA sample. Mix gently by pipetting 10 times without creating bubbles and centrifuge briefly.
5. Perform a PCR using the following parameters according to table 6.
Table 6: Thermal cycler condition for the first PCR
| Lid temperature | Reaction volume | Run time |
| +105 °C | 50 µL | |
| Step | Temperature | Time (hh:mm:ss) |
| 1 | +95 °C | 00:15:00 |
| 2 | +94 °C | 00:00:30 |
| 3 | +60 °C | 00:01:30 |
| 4 | +72 °C | 00:00:30 |
| Go to step 2, repeat X cycles* | ||
| 5 | +68 °C | 00:04:00 |
| Hold | +4 °C | Hold |
* Choose the number of cycles according to the amount of RNA/B cells used as input for the experiment (see table 10).
Clean-up after first PCR
▲ Proceed in the post-PCR workspace.
- Before use, thoroughly vortex the beads until homogenized and no pellet remains at the bottom of the bottle.
- Add 40 μL of beads to each reaction and mix well by pipetting up and down 10 times without creating bubbles.
- Incubate the solution for 5 minutes at room temperature (+15 to +25 °C).
- Place the tubes on the magnetic rack and wait for the solution to clear.
- While the tubes are on the magnetic rack, carefully remove the supernatant.
- Keep the tubes on the magnetic rack and add 200 μL of freshly prepared 80% ethanol. Wait for 30 seconds and remove the ethanol.
- Repeat the washing step with 80% ethanol without removing the tubes from the magnetic rack.
- Remove tubes from the magnetic rack. Briefly spin the tubes down and place them back on the magnetic rack. Remove all residual ethanol carefully, using a small pipette.
- Remove tubes from the magnetic rack. Add 21 μL of Elution Buffer to the beads for elution and mix well by pipetting up and down 15 times without creating bubbles.
- Incubate the solution for 5 minutes at room temperature (+15 to +25 °C).
- Place the tubes on the magnetic rack and wait for the solution to clear.
- Keep tubes on the magnetic rack and carefully transfer 20 μL of the purified first PCR products into new nuclease-free reaction tubes, one for BCR-H and one for BCR-K/L samples.
▲ Note: It is important to remove all the traces of ethanol at this step as it could inhibit subsequent PCR reaction.
▲ Note: The purified PCR product can be stored up to 1 month at −20 °C.
Second PCR amplification
Before start: Thaw BCR-H PCR2 and BCR-K/L PCR2 Primers briefly at room temperature (+15 to +25 °C) and the PCR Master Mix on ice. Mix each component thoroughly by pipetting up and down before use and spin down. Thaw the DNA/RNA UD Indexes (Illumina) briefly at room temperature (+15 to +25 °C) and spin them down. Place all components on ice until use.
1. Prepare primer dilution for BCR-H PCR2-A and BCR-K/L PCR2-B Primers: For this, calculate the volume of PCR2 primers required to process N number of samples (1 μL of diluted primer mix is required per sample). Prepare at least 10 μL. Both primers should be diluted with water by a dilution ratio of 1:10 (e.g., 2 µL primer + 18 µL water = 20 µL of 1:10 primer dilution). Discard the left-over diluted aliquots after use.
2. Prepare two separate PCR2 master mixes in a final volume of 22 μL per reaction without indexes according to tables 7 and 8.
Table 7: BCR-H PCR2 master mix
| Component | Volume (in μL) |
| Water | 8.5 |
| PCR Master Mix | 12.5 |
| BCR-H PCR2 Primers (1:10 diluted) | 1 |
| Total volume per reaction | 22 µL |
Table 8: BCR-K/L PCR2 master mix
| Component | Volume (in μL) |
| Water | 8.5 |
| PCR Master Mix | 12.5 |
| BCR-K/L PCR2 Primers (1:10 diluted) | 1 |
| Total volume per reaction | 22 µL |
3. Mix well by pipetting up and down and centrifuge briefly.
4. For each reaction, add 22 μL of corresponding PCR2 master mix to a nuclease-free PCR tube, respectively, for BCR-H and BCR-K/L samples.
▲ Note: Proceed to the post–PCR workspace.
5. Add 2 μL of the purified first PCR product to each of the BCR-H PCR2 master mix and BCR-K/L PCR2 master mix.
6. Add 1 μL of DNA/RNA UD Index (Illumina) to each reaction. Use a unique index for each reaction and record their assignment.
7. Mix by pipetting up and down 10 times without creating bubbles and centrifuge briefly.
8. Perform a PCR with the BCR-H and BCR-K/L samples using the parameters from table 9.
Table 9: Thermal cycler condition for the second PCR
| Lid temperature | Reaction volume | Run time |
| +105 °C | 25 µL | |
| Step | Temperature | Time (hh:mm:ss) |
| 1 | +95 °C | 00:15:00 |
| 2 | +94 °C | 00:00:30 |
| 3 | +55 °C | 00:01:30 |
| 4 | +72 °C | 00:00:30 |
| Go to step 2, repeat X cycles* | ||
| 5 | +68 °C | 00:04:00 |
| Hold | +4 °C | Hold |
* Choose the number of cycles according to the amount of RNA/B cells used as input for the experiment (see table 10).
▲ Note: The optimal number of second PCR cycles may vary for different templates, cell types, thermal cyclers, and sample amounts. It is recommended to determine the minimal number of PCR cycles required to obtain a sufficient amount (at least 30–40 ng of each library, specific peak, dependent on initial input concentrations and cycles applied) empirically for each experiment (see table 10 & 11).
▲ Note: Given values are based on an adjustment for the 2–3-fold higher BCR-K/L mRNA expression level and the usage of 15 µL eluate for BCR-H and 5 µL for BCR-K/L reaction as input (recommended in section First PCR amplification).
Furthermore, it is recommended to generate parallel libraries of similar nature (e.g., ten samples of 2,000–20,000 sorted B cells each, BCR-H and BCR-K/L chains) using the same number of PCR cycles and mix the obtained libraries in equal volume proportions for sequencing.
This allows obtaining an even coverage in terms of reads-per-cDNA (reads-per-UMI). For example, the library generated with 20,000 B cells may produce more PCR product than one with 2,000 B cells after the same number of PCR cycles. However, the library from 20,000 cells would also carry proportionally more IG cDNA molecules and thus requires more sequencing reads to achieve a comparable coverage.
Table 10: Recommended PCR cycle numbers for sorted memory/naive B cells or PBMCs.
| Estimated count of memory/naive B cells in a sample | Amount of template RNA purified from sorted B cells or PBMC | Recommended number of first PCR cycles (total cycles) | Recommended number of second PCR cycles (total cycles) | Recommended number of sequencing reads |
| 1×102 – 5×102 | 0.1–5 ng | 25 | 13–16 | 100 reads per input memory/naive B cell (calculated per library based on total input amount) |
| 5×102 – 5×103 | 0.5–50 ng | 21 | 12–15 | |
| 5×103 – 5×104 | 5–500 ng | 18 | 12–15 | |
| 5×104 – 5×105 | 50–500 ng | 18 | 10–13 |
Overview of recommended total PCR cycle numbers for sorted memory/naive B cells or PBMC.
▲ Note: Consider that plasma cells may prominently increase in PBMC after infection or vaccination. Keep in mind that plasma cells may have 100– to 1,000-fold higher levels of IG mRNA expression, which requires appropriate sequencing coverage as described in table 11. For average PBMC samples, please refer table 12 for recommended sequencing depth.
Table 11: Recommended PCR cycle numbers for sorted plasma cells and plasmablasts. Note that plasma cells express 100– to 1,000-fold more IG mRNA compared to other B cell populations.
| Number of plasma cells in a sample | Recommended number of first PCR cycles (total cycles) | Recommended number of second PCR cycles (total cycles) | Recommended number of sequencing reads |
| 10–200 plasma cells | 22 | 12–15 | 25,000 reads per input plasma cell |
| 200–2,000 plasma cells | 20 | 12–15 |
▲ Note: The given number of sequencing reads represents the recommended coverage that allows for UMI‑based elimination of amplification biases and accumulated errors.
Quality control and pooling
- Verify quality and concentration of the obtained PCR product by capillary electrophoresis (TapeStation® or Bioanalyzer®).
The fragment size of the library should have a peak at approximately 530–600 bp for BCR-H and 400–480 bp for BCR-K/L chains respectively, when using D5000 ScreenTapes (TapeStation). Please note that using other screen tapes or BioAnalyzer reagents, the peak range can differ. - If multiple libraries were generated, pool the obtained second PCR libraries of similar nature (i.e. libraries generated with similar cell/RNA input amounts) separately for BCR-H and BCR-K/L. To do this, combine equal volume portions from each individual sample after the second PCR.
If 1–4 libraries are prepared at the same time, the whole reaction volumes of the second PCRs can be pooled. For more than four samples, pool 5–10 µL from each second PCR. The total amount of pooled DNA products should be at least 50 ng.
Library clean-up
Before start: Adjust beads to room temperature (+15 to +25 °C) according to manufacturer’s instructions.
Purify the obtained pooled DNA using bead protocol with 1:0.8 sample-to-bead ratio as follows:
- Before use, thoroughly vortex the beads until homogenized and no pellet remains at the bottom of the tube.
- Аdd the volume of beads equal to 0.8 the volume of pooled DNA to each reaction and mix well by pipetting up and down 10 times without creating bubbles. Incubate the solution for 5 minutes at room temperature (+15 to +25 °C).
- Place the tubes on the magnetic rack and wait for the solution to clear.
- While the tubes are on the magnetic rack, carefully remove and discard the supernatant.
- Add 200 µL of freshly prepared 80% ethanol to the tubes, while on the magnetic rack. Wait for 30 seconds and remove the ethanol.
- Repeat the washing step with 80% ethanol without removing the tubes of the magnetic rack. Remove tubes from the magnetic rack. Briefly spin the tubes down and place them back on the magnetic rack. Remove all the residual ethanol carefully, using a small pipette.
▲ Note: It is important to remove all traces of ethanol at this step as it could inhibit subsequent sequencing procedures. - Remove tubes from the magnetic rack. Add 21 μL of Elution Buffer for elution and mix well by pipetting up and down without creating bubbles.
- Incubate the solution for 5 minutes at room temperature (+15 to +25 °C).
- Place the tubes on the magnetic rack and wait for the solution to clear.
- Keep tubes on the magnetic rack and carefully transfer 20 μL of purified second PCR products into a new tube, respectively for BCR-H and BCR-K/L samples.
▲ Note: When generating libraries with low RNA input amounts, a non-specific band (smaller than target band) can be observed. If this is the case, please perform an additional clean-up (after the final 0.8× library clean-up) with a modified bead-to-sample ratio of 0.7× (see section Library clean-up) to remove/reduce the unspecific band.
Final quality control
- Verify the quality of the pooled BCR-H and BCR-K/L libraries by analyzing an aliquot by capillary electrophoresis (Bioanalyzer or TapeStation) (see figure 3 for an example of Agilent TapeStation traces). Quantify the obtained purified pools using Qubit fluorimeter and Qubit dsDNA High Sensitivity Assay Kit.
- BCR-H and BCR-K/L pools can be mixed in equimolar ratio for sequencing, considering that initially ¾ of cDNA was used for BCR-H and ¼ for BCR-K/L library preparations.
▲ Note: Purified PCR product can be stored for 1–3 months at −20 °C.
The obtained purified pooled libraries are ready for Illumina sequencing. Before sequencing, libraries must be quantified. To achieve the most accurate results and ensure proper loading of the sequencing flow cell, it is recommended to use a qPCR-based quantification method. Electropherogram molarity and/or Qubit quantification are not recommended.
Please find the initial recommendations below, which can be used as a starting point:
MiSeq (not MiSeq i100)
Library loading: Start with a low library concentration of, for example, 10 pM. Check the sequencing quality metrics, such as cluster density, the percentage of passing filter and the percentage of reads aligned to PhiX. If necessary, adjust the loading concentration stepwise, in line with Illumina’s recommendations.
PhiX spike-in: 20%.
NextSeq 2000
Library loading: Start with a low library concentration of, for example, 650 pM (on-deck denaturation), depending on the cartridge size and chemistry (Standard or XLEAP). Check the sequencing quality metrics, especially the percentage of passing filter, percentage of occupancy and the percentage of reads aligned to PhiX. If necessary, adjust the loading concentration stepwise in line with Illumina’s recommendations.
PhiX spike-in: 20%.
To achieve reliable UMI-based correction of amplification bias and sequencing errors, target a sequencing depth of 100 sequencing reads per input naive/memory B cell. For plasma cells and plasmablasts target 25,000 sequencing reads per input cell. For average PBMC samples please refer to table 12.
This should be calculated for each library based on the total input used during cDNA synthesis. We recommend targeting the following coverage metric for sequencing data analysis: average reads per UMI > 30.
Pooled libraries should be analyzed using at least 250 × 250 bp paired-end Illumina sequencing with standard sequencing primers. If possible, 300 × 300 bp paired-end Illumina sequencing is recommended. To avoid any possible discrepancies between separate sequencing runs and minimize batch effects, it is recommended, if possible, to sequence all samples that will be compared in downstream analysis within the same run to ensure that comparative analyses remain unbiased.
Table 12: Recommended sequencing depth and run parameters.
| Sequencing depth | Suggestion for average PBMC samples: minimum 800 reads per input B cell (calculated for each library based on the total input used during cDNA synthesis).* For sorted cells: 100 reads per input naive/memory B cell 25,000 reads per input plasma cell. | |
| Sequencing type | Paired-end, Dual indexing | |
| Sequencing read Read 1 – i7 – i5 – Read2 | MiSeq/NextSeq 2000 | 250 – 10 – 10 – 250 |
* When starting with RNA as input, approximately 1 ng of total RNA corresponds to ~1000 cells.
For other sequencing platforms, these parameters need to be determined empirically following Illumina’s guidelines for low-diversity libraries.
For advanced analysis, BCR libraries generated with the BCR RNA Profiling Kit, human are fully compatible with both MiLaboratories analysis tools:
MiXCR is a widely adopted tool for comprehensive BCR repertoire analysis and Platforma is a no-code bioinformatics platform for NGS data analysis. MiXCR enables accurate alignment, clonotype assembly, and annotation of sequencing data, providing high-confidence insights into B cell diversity and immune responses. Platforma enables powerful downstream analyses including repertoire diversity assessment, gene usage, rarefaction, and data visualization, without requiring any coding expertise.
Get started with either MiXCR as a stand-alone command line tool at https://mixcr.com/mixcr/getting-started/installation or use it as a block in Platforma’s graphical user interface at https://platforma.bio/downloads. Choose the preset protocol: BCR RNA Profiling Kit, human in the Miltenyi Clonotyping block.
If you have questions about the quality of your library quality control, please contact Miltenyi Biotec Technical Support.
Low BCR-H and/or BCR-K/L PCR product yield or no BCR-H and/or BCR-K/L product
Reason: RNA contains impurities that inhibit cDNA synthesis
Solution: In some cases, ethanol precipitation or additional column-based purification of RNA can remove impurities. If this does not help, re-isolate the RNA.
Reason: RNA is heavily degraded.
Solution: Re-isolate the RNA. To lower the risk of RNA degradation and contamination general recommendations should be implemented, such as using gloves, tips with aerosol filters, and certified RNase-free reagents.
Reason: Decreased cDNA synthesis/yield
Solution 1: Do not use heparin-coated tubes for blood collection. Heparin dramatically decreases cDNA synthesis efficiency.
Solution 2: During the elution steps, beads clumping may significantly reduce the cDNA yield, which is critical at the cDNA purification step. To maximize the cDNA yield, it is recommended to break up the clumps using pipetting and heating at +50 °C for 10–30 minutes until completely dissolving.
Solution 3: During the clean-up steps, it is important to remove all traces of ethanol, as residual ethanol can inhibit subsequent PCR reaction.
Reason: PCR undercycling
Solution: Repeat the second PCR amplification using two or three more PCR cycles. If this does not help, repeat the cDNA synthesis reaction and the first PCR reaction using three more PCR cycles.
BCR-H and/or BCR-K/L bands and the background smear are very intense in the TapeStation/ Bioanalyzer image
Reason: PCR overcycling
Solution: Repeat the second PCR amplification, using two or three fewer PCR cycles than in the previous experiment.
BCR-H and/or BCR-K/L bands are present, but the background smear is intense or alternative length fragments are visible in the TapeStation/Bioanalyzer image
Reason: Low IG RNA content in the initial RNA sample
Solution: Purify target library using beads with size selection protocol or agarose gel purification. For removal of smaller (than target) non-specific bands observed in TapeStation/Bioanalyzer image, please either modify beads-to-sample ratio in the final clean-up in order to remove/reduce these unspecific bands or increase sequencing depth.
BCR cDNA libraries can be generated starting from RNA isolated from sorted/isolated B cells, PBMCs, or any B cell–containing tissues. Control for the counts of B cells of interest in a sample is desirable for downstream data analysis.
It is preferable to use freshly isolated cell samples. For frozen samples, it is recommended to culture thawed cells overnight in RPMI 1640 supplemented with 10 % human serum.
If using freshly isolated cells is not an option, lysed cells/tissues can be stored in RLT buffer (QIAGEN) at –70 °C for at least 6 months.
Larger samples (>50,000 cells) can be also stored in TRIzolTM reagent at –70 °C for up to 6 months.
For RNA isolation, RNA purification kits or TRIzolTM reagent are recommended. When choosing a purification kit, please make sure that it is appropriate for your sample type, input amount, and includes DNase treatment. The last point is critical since genomic DNA can significantly affect cDNA synthesis and subsequent PCR amplification. Therefore, it is strongly recommended to include DNase treatment in the RNA extraction procedure or perform independent DNase digestion with additional purification of RNA before starting the experiment. For large numbers of cells (> 200,000 cells) on-column DNase treatment protocols can be used without modification. For small samples (<200,000 cells) it is recommended to dilute DNase I stock solution 4 times before treatment to decrease losses. Moreover, using uniform conditions for all samples in the experiment is advisable to ensure that comparative analyses remain unbiased.
For large samples (>50,000 cells)
Verify quantity and quality of extracted RNA, using Qubit and Agilent Bioanalyzer or TapeStation systems. RNA integrity number >7, or correct 28S rRNA:18S rRNA ratio (around 1.5–2.5:1) and a low number of shadow bands above and below 18S band are indicative of high-quality RNA.
For small samples (i.e. <50,000 cells)
To avoid losing precious sample it is not mandatory to check quantity and quality of extracted RNA for small samples. For small RNA amounts, it is preferable to start cDNA synthesis immediately after RNA extraction.
Carrier poly(A) RNA can be used during the RNA extraction procedure to increase the yield of isolated RNA. Carrier RNA will not interfere with primers for cDNA synthesis reaction.
Recommendations to choose the appropriate strategy for RNA isolation
1. PBMCs
Do not use heparin coated tubes for blood collection. Heparin dramatically decreases cDNA synthesis efficiency.
Perform isolation of mononuclear cells from whole blood. If done by density gradient centrifugation, spin down cells at 350 × g for 15 minutes. Remove supernatant.
Add at least 50 μL (or 50 μL per 2 × 105 cells) of RLT buffer and mix by pipetting up and down without creating bubbles. The samples can either be stored at –70 °C for up to 6 months, or directly processed for RNA purification using RNA isolation kits, or RNA purification using 4 volumes of TRIzolTM reagent added per 1 volume of RLT buffer (see appendix B).
Alternatively, place cells in at least 300 μL (or 300 μL per 3×106 cells) of TRIzolTM reagent and extract RNA using TRIzol reagent protocol (see appendix B).
2. From 100 to 50,000 sorted or purified B cells
Sort/place cells directly in 50–300 μL RLT buffer. The volume of RLT buffer should not be diluted more than 20% during sorting.When using a 70 µm nozzle for cell sorting, 1,000 sorted cells may carry a volume of around 1 µL. The cells are lysed immediately in the collection tube and mRNA is protected from degradation.
Lysed cells can be stored in RLT buffer at –70 °C for at least 6 months.
Use RNA isolation kits for RNA purification.
To obtain the maximum output for small samples (100–10,000 cells), add 4 volumes of TRIzolTM reagent to RLT cellular lysate and extract RNA using TRIzolTM reagent protocol (see appendix B).
The minimal number of cells for input into the protocol is 100 B cells or 10 plasma cells sorted directly into RLT buffer, followed by RNA extraction with TRIzolTM reagent (see appendix B). To isolate RNA from comparable cell numbers, use co-precipitant and carrier RNA (or both, depending on isolation method) to avoid losses.
3. More than 50,000 sorted or purified B cells
Sort/place cells into 300 μL of PBS, then spin down the cells at 350 × g for 15 minutes. Remove supernatant.
Add at least 50 μL (or 50 μL per 2 × 105 cells) of RLT buffer and mix by pipetting up and down without creating bubbles. The samples can either be stored at –70 °C for up to 6 months, or directly processed for RNA purification using RNA isolation kits, or RNA purification using 4 volumes of TRIzolTM reagent added per 1 volume of RLT buffer (see appendix B).
Alternatively, place cells in at least 300 μL (or 300 μL per 3×106 cells) of TRIzolTM reagent and extract RNA using TRIzolTM reagent protocol (see appendix B).
4. B cell–containing tissue
For larger tissue samples, it is recommended to dissociate the sample before to ensure uniform RNA extraction. Ideally, obtain a single cell suspension using incubation with DNase and proteases mixes. Optionally, wash cells with PBS. Immediately proceed with RNA isolation kits using RLT buffer or use TRIzolTM reagent (see appendix B) for RNA extraction.
- Lyse cell pellet in TRIzolTM Reagent by repetitive pipetting. Use at least 300 µL TRIzolTM Reagent per 100 – 3×106 cells. For larger cell amounts, increase the volume of the reagent according to the proportion of 1 mL per 5–10×106 cells. For small samples (<50,000 cells), carrier poly(A) RNA can be added directly to the lysate.
- Alternatively, to step 1: If the cell lysate is in RLT buffer, add 4 volumes of TRIzolTM Reagent and mix well by vortexing.
- Incubate the homogenized samples for 5 minutes at room temperature (+15 to +25 °C). Add 1/5 volume of chloroform and mix well by vortexing.
- Incubate the tubes at room temperature (+15 to +25 °C) for 3 minutes.
- Centrifuge the samples at >10,000 × g for 10 minutes at +2 to +8 °C. After centrifugation, put samples on ice. The mixture separates into a lower red, phenol-chloroform phase, an interphase, and a colorless upper aqueous phase. The RNA remains exclusively in the aqueous phase.
- Transfer the colorless aqueous phase to a fresh tube without disturbing the interphase. Add 1 μL of co-precipitant (Pellet Paint® #70748 Merck Millipore or analog) and mix well by pipetting.
- Precipitate the RNA from the aqueous phase by mixing with 0.5 mL of isopropanol per 1 mL of TRIzolTM Reagent used for lysis. Incubate samples at +15 to +30 °C for 10 minutes.
- Centrifuge the tubes at >10,000 × g for 10 minutes at +2 to +8 °C. The RNA forms a visible pellet on the bottom of the tube.
- Remove the supernatant without disturbing the pellet.
If starting from <50,000 cells
- Add 1 mL of freshly prepared 75% ethanol to the RNA pellet. Mix the sample by vortexing.
- Centrifuge the tubes at >10,000 × g for 5 minutes at +2 to +8 °C. Remove supernatant.
- Dry the RNA pellet completely but be careful to not overdry the pellet as this will decrease its solubility.
- Dissolve RNA in 10 μL of RNase-free water and immediately proceed to cDNA synthesis.
If starting from >50,000 cells
- Add 1 mL of freshly prepared 75% ethanol to the RNA pellet. Mix the sample by vortexing.
- Centrifuge at >10,000 × g for 5 minutes at +2 to +8 °C. Remove supernatant.
- Dry the RNA pellet completely but be careful to not overdry the pellet as this will decrease its solubility. Dissolve RNA in 10 μL of RNase-free water.
- Proceed to cDNA synthesis or store RNA at –70 °C.
1 product found
|