Immunostaining and clearing of organoids for 3D imaging using MACS® UltraMount 48 - immunostaining after tissue embedding


This application protocol describes step-by-step the immunostaining after tissue embedding and clearing of organoids and spheroids using the MACS UltraMount 48 and the MACS Clearing Kit for downstream 3D imaging on the UltraMicroscope Blaze™ Light Sheet System. The MACS Clearing Kit enables the immunostaining and clearing of up to 4×48 samples of <2 mm in diameter.

▲ Note: Preliminary assessment suggests that the MACS® Deep Clearing Kit (order no. 130-136-470) is compatible with this protocol, when using the corresponding solutions and incubation times. For better results, user-side optimization may be required.

Please use Deep Antibody Staining Solution (8x) instead of 10x Antibody Staining Solution, Deep Permeabilization Solution instead of Permeabilization Solution, and Deep Clearing Solution instead of Clearing Solution.

Protocol

  • Always freshly prepare reagents. Failure to do so may lead to suboptimal results.

Preparation of PFA buffer

  • All steps that include paraformaldehyde (PFA) must be performed under a fume hood.
  • The protocol describes the preparation of 500 mL of 4% PFA buffer. For higher or lower volumes scale the amounts of the components accordingly.
  • Do not use ready-to-use PFA solutions.

  • Add 20 g of PFA powder to 400 mL of PBS and stir the mixture at +60 °C under a fume hood until it is dissolved.
    ▲ Note: Do not boil!.
    ▲ Note: PFA powder does not dissolve instantly.
  • Add 0.2–0.4 g of NaOH and continue stirring at +60 °C until the solution clears up. This may take 2–3 hours.
  • Adjust the pH to 7.2–7.4 with HCl and NaOH.
  • Adjust the volume to 500 mL with PBS.
  • Filter the solution using a 0.22 µm filter.
  • PFA buffer can be stored at +4 °C for one week. For long-term storage prepare aliquots and store at –20 °C. Avoid repeated freeze/thaw cycles.

Preparation of 1× Antibody Staining Solution of the MACS® Clearing Kit

  • If precipitation occurs due to storage of the 10× Antibody Staining Solution at low temperature, mix well before preparation of 1× Antibody Staining Solution.
  • 10× Antibody Staining Solution from the MACS Clearing Kit needs to be diluted 1:10 with sterile distilled water before use, e.g., by diluting 1 mL of 10× Antibody Staining Solution with 9 mL of sterile distilled water. Adjust the volume of 1× Antibody Staining Solution according to the number of samples to be cleared.
    ▲ Note: When using MACS Deep Clearing Kit, Deep Antibody Staining Solution (8×) needs to be diluted 1:8 with PBS before use. Prepare Deep Antibody Staining Solution (1×) by diluting 1 mL of Deep Antibody Staining Solution (8×) with 7 mL of PBS. Adjust the volume of Deep Antibody Staining Solution (1×) to the amount of tissue to be cleared.
  • Always use freshly prepared 1× Antibody Staining Solution.

Preparation of ethanol/Tween® 20 dilutions for dehydration

  • The following solutions are needed:
    • Ethanol 50%/Tween® 20
    • Ethanol 70%/Tween® 20
    • Absolute (≥99.8%) ethanol/Tween® 20
  • Prepare ethanol dilutions of 50% and 70% in water. Use absolute ethanol (≥99.8%) and sterile distilled water to prepare the dilutions. Add Tween® 20 to each dilution for a final concentration of 2%.
    ▲ Note: Make sure to use fresh absolute ethanol (≥99.8%) that does not contain water residues.

Preparation of Clearing Solution from the MACS Clearing Kit

  • Clearing Solution is ready-to-use, but may freeze when stored at +2 to +8 °C. Therefore, it is recommended to prepare aliquots and store at +2 to +8 °C. Bring aliquots to room temperature (+20 to +25 °C) before proceeding to clearing of samples.
    ▲ Note:
    Use polypropylene tubes for preparation of aliquots. Other plastics might not be compatible with the Clearing Solution.

Shop required products for this application protocol

4 products found

|

You have previously chosen International as your location.
What do you want to do?