Miltenyi Biotec
Miltenyi Biotec home - back to start page
Miltenyi Biotec
  • MACS® Cell Analysis
Miltenyi Biotec
Miltenyi Biotec
Miltenyi Biotec
Miltenyi Biotec
Contact technical support Contact technical support
Instrument Service Instrument Service
MACS Academy MACS Academy
Miltenyi TV Miltenyi TV
MACS® Technology MACS® Technology
FAQs FAQs
  MACS® Technology for cell separation
  MACS® Columns for cell separation
  Titration of antibodies for cell separation with MACS® MicroBeads for indirect magnetic labeling
  Isolation of hematopoietic progenitor cells
  Dendritic cell isolation
  Tumor cell enrichment and detection
  Isolation of mouse cells
  MACSelect™ Transfected Cell Selection
  mRNA isolation and cDNA synthesis
  MACSQuant® Analyzer
Forum Forum
Publications Publications
New products New products
Data sheets & protocols Data sheets & protocols
Certificate of Analysis Certificate of Analysis
Miltenyi Biotec
MACS eNews
Subscribe to our
MACS eNews
for regular product information via e-mail.
Miltenyi Biotec
Questions regarding: MACSQuant® Analyzer
Which lasers are used in the MACSQuant® Analyzer optical bench?
What are the fluorescence sensitivities of the MACSQuant® Analyzer?
What is the error in absolute cell count?
What is the dynamic range of sensitivity?
What type of sample tubes can I use?
How many events can be acquired on the MACSQuant® Analyzer?
At what speed can the MACSQuant® Analyzer analyze samples?
What is the minimum sample volume that can be analyzed by the MACSQuant Analyzer?
What are the dimensions of the flow cell?
What is the laser spot for the MACSQuant® Analyzer?
What is the carryover between samples?
What are the specifications of the workstation?
What are the dimensions of the MACSQuant® Analyzer?
What are the optimal operating conditions?
What is the power usage of the MACSQuant ®Analyzer?
What type of sheath fluid should I use?
Does Miltenyi Biotec offer flow cytometry analysis software?
How many logarithmic decades does the MACSQuantify™ Software display?
What is the signal resolution?
Does the MACSQuant® Analyzer display data in bi-exponential or hyperlog scale?
Questions & answers
Which lasers are used in the MACSQuant® Analyzer optical bench?
Violet (405nm) Blue (488nm - diode pumped solid state) Red (635nm)

What are the fluorescence sensitivities of the MACSQuant® Analyzer?
MESF: FITC <200; PE <100; APC <150

What is the error in absolute cell count?
The error in the absolute cell count is less than 5% CV

What is the dynamic range of sensitivity?
The dynamic Range for FSC/SSC is 1 µm – 15 µm

What type of sample tubes can I use?
The MACSQuant® Analyzer with the single tube holder or the MACS® MiniSampler can process many types of sample tubes. Microcentrifuge tubes, 5-, 10-, 15- mL tubes and 96-well microtiter plates can be automatically processed by the MACSQuant Analyzer.

How many events can be acquired on the MACSQuant® Analyzer?
The MACSQuant can acquire up to 10,000 events per second

At what speed can the MACSQuant® Analyzer analyze samples?
The MACSQuant Analyzer has three different flow rates Low (25uL/min), Med (50uL/min) and High (100 uL/min)

What is the minimum sample volume that can be analyzed by the MACSQuant Analyzer?
25 µL.

What are the dimensions of the flow cell?
The flow cell is 200 × 250 µm and has a rectangular profile

What is the laser spot for the MACSQuant® Analyzer?
The Laser Spot is 15 × 45 µm

What is the carryover between samples?
The carryover between samples on the MACSQuant Analyzer is <0.01%

What are the specifications of the workstation?
The MACSQuant® Analyzer workstation includes:

PC with intel Pentium M 1.8 GHz
Memory 1 GB RAM
Data Storage: 40GB hard drive DVD RW drive
Networking: Ethernet
Monitor: 12 inch, touch screen
2x USB port Analog RGB / VGA port
COM Port (DSUB-9)

What are the dimensions of the MACSQuant® Analyzer?
The Size (W × D × H) of the MACSQuant is 600 × 350 × 400 mm

What are the optimal operating conditions?
The optimal operating temperature is between 15 to 25°C with humidity ranging from 10 to 90%.

What is the power usage of the MACSQuant ®Analyzer?
The power is 350 W

What type of sheath fluid should I use?
Miltenyi Biotec has specially designed buffers

Does Miltenyi Biotec offer flow cytometry analysis software?
Yes, MACSQuantify™ Software is a user-friendly software that can be used with data acquired from the MACSQuant® Analyzer and other flow cytometers (fcs2.0 and fcs3.0 file formats)

How many logarithmic decades does the MACSQuantify™ Software display?
The MACSQuantify Software displays 5-decade logarithmic scales

What is the signal resolution?
The signal resolution of the MACSQuant® Analyzer is 16bit (65,536 channels) on all parameters

Does the MACSQuant® Analyzer display data in bi-exponential or hyperlog scale?
The MACSQuantify™Software, which runs on the MACSQuant® Analyzer, utilizes a new version of scale called hlog, which is similar to bi-exponential and hyperlog scales.
FAQ
MACS® Technology for cell separation (5)
MACS® Columns for cell separation (7)
Titration of antibodies for cell separation with MACS® MicroBeads for indirect magnetic labeling (4)
Isolation of hematopoietic progenitor cells (4)
Dendritic cell isolation (5)
Tumor cell enrichment and detection (3)
Isolation of mouse cells (4)
MACSelect™ Transfected Cell Selection (2)
mRNA isolation and cDNA synthesis (6)
MACSQuant® Analyzer (20)
Miltenyi Biotec
Search this website
Advanced search
Toolbox
Customer login
Contact
Newsletter
Help
Favorites / Prices
Request prices or add products to your favorite list.
Details
© 1999 - 2012 Miltenyi Biotec ¦ Terms & conditions ¦ Legal notes
Miltenyi Biotec website